Enter the sensor information fields. Use the F1 function key (
Confirm your choice using the VAL key.
To add a value in the table of calibration points, use the
value and value read) then confirm using the VAL key.
Repeat this operation for all the calibration points (maximum of 4).
To delete a line, select it then use the
key.
TM612
) to move from one field to the next.
keys, enter the calibration points (real
- 19 -
TM612
To edit a line, select it then use the navigation key ( ) to make editing possible.
Confirm using the VAL key to return to the measurement screen.
To ensure the measurements are made using the calibration coefficients defined earlier, go to the
Configuration/Function menu.
In the Probe field, select sensor1 (SONDE1-SENSOR1 below).
Note: the Calibrated sensors are at the top of the list and their name is preceded by an *.
Confirm the latter using the VAL key.
The chosen calibrated sensor is displayed in the measurement screen.
Name of
calibrated sensor
used
D.4

Storage of acquisitions in progress.

The TM612 is designed to store 10,000 values in one or more acquisition bursts.
Using the F2 key, enter the Mode menu.
Select the Memory function.
Confirm using the VAL key.
- 20 -
- 19 -