1 Installation At delivery F2 is delivered in the transport mode. This means that only the real time clock is active. No measurements take place in this position. The transport mode is indicated by a "no" in the upper left hand corner.
Connections The screw terminal blocks are situated under the cover on the rear side of the calculator. Note: The covers retaining screws is protected by a labelled sea. When the calculator is mains powered the will be a permanent power cord. Note: BR431 power supplied, flow sensor has to be connected to +3V.
2 Mounting F2 can be mounted either on the flow sensor or wall mounted. When the calculator is mounted on the flow sensor, the adapter provided for this purpose has to be used. Mounting on a flow sensor The adapter allows the calculator to be mounted vertically or horizontally, see fig.
Installation test When installation has been completed simple test can be done to verify that the calculator has been installed correctly. Wait until the flow sensor supplies a pulse. This shall result in the symbol for flow sensor pulse, a square shown on display will flashing once.
3 Handling Display The F2 has an LCD (Liquid Crystal Display), see fig. 3.1a below. The values are displayed in sequences. The left-hand sequence digit shows the relevant sequence. The right-hand sequence digit shows which value in the sequence is being displayed. A short button depression results in a move to the next value in the relevant sequence.
Display sequence table Description Accumulated energy (Default position) Accumulated volume according to flow sensor1 Display test Accumulated volume for pulse input 1 (Only when F2 fitted with pulse inputs) Accumulated volume for pulse input 2 (Only when F2 fitted with pulse inputs) Error code, see Error code Error time, [Minutes]...
Error codes Error codes are shown in display sequence ”15”. Code Description 0001 Break in low temperature sensor (L) 0002 Short circuit in low temperature sensor (L) 0004 Break in high temperature sensor (H) 0008 Short circuit in high temperature sensor (H) 0005 Error code combination 0001 + 0004 000A...
4 Service Certain parameters in F2 can be altered using the display button when F2 is set in the service mode. Procedure to enter the service sequence: 1. Hold the service button on the rear side of the calculator pressed by using a small screwdriver.
0 = Installed at low end (L) 1 = Installed at high end (H) Recommended battery replacement date [YYMMDD] Do not change without consulting with Metrima AB. Exit service sequence, Fig. 4.1b, Test key 1 = EXIT 2 = return to service sequence ”00”...
5 Verifying the calculator Verification of the calculator's measurement accuracy is undertaken in the test mode, where the energy value/flow sensor pulse is issued via the HF-output at the service adapter. For each flow sensor pulse, measurement takes place on the temperature sensors and a pulse burst corresponding to the measured energy of the meter is issued.
7 Technical data Power supply Battery 3V – 2,2Ah Operation time max. 10 years Mains 230V±10%, 45-65Hz Table 7.1, Power supply Temperature sensors Approved and matching pares of temperature sensors of the type Pt 100 or Pt 500 are to be used.
Flow sensor placing F2 can be configured for flow sensor placement in high or low end of the pipe (supply or return pipe). This is marked H = high or L = low in display sequence “64”. Figure 7.6, Flow sensor placing Maximum values for power The values below are valid for energy unit [MWh] and standard decimal setting Pulse value...
7.10 Pulse outputs (either pulse outputs or pulse inputs on F2) F2 can be equipped with two pulse outputs as standard of the type “Open collector” for energy (pulse output 1) and volume (pulse output 2). The last digit and the energy unit will reveal the pulse weight of the pulses. Example: sequence “10”...
Article number F2 F2 ABCDEFGHIJ KLM Pt100 2-wire measurement, flow in low temperature Pt100 2-wire measurement, flow in high temperature Pt500 2-wire measurement, flow in low temperature Pt500 2-wire measurement, flow in high temperature Battery supply Bus supply Mains supply 230VAC 24 VAC Kt Input 2.5 l/p Kt Input 25 l/p...
Data output table The following data are accessible via the data output: SIOX Manufacturers Data specific 60870-5 (option) Flow sensor placing Program version Manufacturer Communication address Meter number Error code (limited) Accumulated energy Accumulated, volume 11 Accumulated, volume 22 Flow temperature (high) Return temperature (low) Temperature difference “Operation...