IF FNPs _ err < >0 THEN Finish
490
!
500
OUTPUT @Ps;"VSET ";Chan,"0 ;ISET '';Chan,"0"
510
!
520
INPUT "ANY MORE OUTPUTS TO CALIBRATE? (Y OR N)",X$
530
IF (X$=''Y" OR X$=''y") THEN Start_loop
540
!
550
OUTPUT @Ps;"CMODE 0"
560
!
570 Finish: ! HERE WHEN DONE
580
OUTPUT @Ps;''CLR"
590
DISP "END OF CALIBRATION PROGRAM"
600
END
610
!
620
!
630
DEF FNDvm
640
COM /Instr/ (@Ps,@Vm
650
WAIT.02
660
TRIGGER @Vm
670
ENTER @Vm;Reading
680
RETURN Reading
690
FNEND
700
!
710
DEF FNPs _ err
720
COM /Instr/ @Ps,@Vrn
730
OUTPUT @Ps;''ERR?"
740
ENTER @Ps;Err
750
IF Err< >0 THEN PRINT "POWER SUPPLY ERROR # '';Err;" CORRECTION CONSTANTS NOT SAVED -
RESTART. "
760
RETURN Err
770
FNEND
LINE 10: This comment line identifies the program as a CALIBRATION EXAMPLE.
LINE 30,40: Assigns I/O path names to the power supply and the voltmeter.
LINE 50: Establishes a COM block for the instruments on the GP-IB.
LINE 60: Initializes the variable Shunt _ resistor to .1 ohms.
LINE 80: Clears the power supply.
LINE 90: Initializes the voltmeter to take voltage readings.
LINE 100,110: Tells the voltmeter to take a reading and clears the voltmeters output buffer. This reading is not used in
the program.
LINE 120: Turns on the power supply calibration mode.
LINE 140: Labels the line ''Start_loop" to loop back to when calibrating more than one output.
LINE 160: Enters the output channel number to be calibrated.
LINE 170,180: Prompts the user to make voltage calibration connections and waits for CONTINUE key to be pressed.
LINE 200: Sets the voltage of the specified output to the high calibration point.
LINE 210: Sets the variable Vhi to the output voltage as measured by the user defined function FNDvm.
LINE 230: Sets the voltage of the specified output to the low calibration point.
LINE 240: Sets the variable Vlo to the output voltage as measured by the user defined function FNDvm.
LINE 260: Sends the measured data in volts to the power supply for the specified channel.
LINE 280: Checks the power supply for errors with the user defined function FNPs _ err. If there was an error, the program
goes to the line labeled Finish and the supply is cleared. The new measured data is not used and the previous
calibration constants are maintained.
LINE 300: Instructs the power supply to perform an overvoltage calibration on the specified channel.
LINE 310--330: Displays a message on the computer until bit 4 (RDY) of the power supply's serial poll register indicates
that the supply is finished processing the OVCAL command. This may take up to 10 seconds.
LINE 350: Checks for errors. See line 280.