Collection Windows

collection windows allow for collecting only the effluent
desired and discarding the rest. optionally, build up to ten
collection windows.
collection windows are useful for:
• excluding void volumes
• excluding equilibration or washing volumes
• aiding in selective peak collection
Remember four key points about collection windows:
• it is always possible to collect from start to end,
designating the entire collection period as one
collection window.
• Whenever the fraction collector is outside a collection
window, it automatically diverts the effluent to drain.
• At the start of each collection window, the dispense
head automatically advances to the next tube.
• The collection windows are not mode specific.
Windows created in one mode are active in all other
modes.