System Specifications; Software - Agilent Technologies 34970A Manual

Hide thumbs Also See for 34970A:
Table of Contents

Advertisement

System Specifications

Scanning Inputs
Analog
34901A, 34902A, and 34908A
multiplexer channels
Digital
34907A digital in and totalize
Scan list
Scans channels in ascending order
Scan Triggering
Source
Interval, external, button press, software,
or on monitor channel alarm
Scan count
1 to 50,000 or continuous
Scan interval
0 to 99 hours; 1ms step size
Channel delay
0 to 60 seconds per channel; 1 ms step size
External trig delay
<300 µs. With monitor on <200 ms
External trig jitter
<2 ms
Alarms
Analog inputs
Hi, Lo, or Hi + Lo evaluated each scan
Digital inputs
34907A digital in maskable pattern match
or state change
34907A totalize: Hi limit only
Monitor channel
Alarm evaluated each reading
Alarm Outputs
4 TTL compatible
Selectable TTL logic Hi or Lo on fail
Latency
5 ms (typical)
Memory
Battery backed, 4 year typical life
Readings
50,000 with timestamp
Readable during scan
States
5 instrument states with user label
Alarm Queue
Up to 20 events with channel number,
reading, and timestamp
System Features
Per-channel Math
Individual Mx + B scaling and
Min/Max/Average calculated real time
Power Fail Recovery
Resumes scanning automatically
Relay maintenance
Counts each relay closure and stores
on module User resettable
Real-time clock
Battery-backed, 4-year typical life
General Specifications
Power Supply
100V/120V/220V/240V ±10%
Power Line Frequency
45 Hz to 66 Hz automatically sensed
Power Consumption
12 W (25 VA peak)
Operating Environment
Full accuracy for 0°C to 55°C
Full accuracy to 80% R.H. at 40°C
Storage Environment
-40°C to 70°C
Weight
Net: 3.6 kg (8.0 lbs)
Safety
Conforms to CSA, UL-1244, IEC 1010 Cat I
RFI and ESD
CISPR 11, IEC 801/2/3/4
16
[1]
[1]
[1]

Software

Agilent BenchLink Data Logger II
(not included with Option 001)
System Requirements
[2]
Operating System
Windows 98SE, NT
Adobe
(to view documentation)
Microsoft
(required when using Windows NT)
Controller
Recommend Pentium
greater, Min: Pentium III, 500 MHz
RAM
Recommend 256MB or greater, Min 128MB
Disk Space
Recommend 200MB, Min 70MB
Display
800x600 resolution, 256 colors
Computer Interfaces
[3]
GPIB
Agilent and National Instruments PCI-GPIB
LAN-to-GPIB
E5810A
USB-to-GPIB
82357A
RS-232 (Serial Port) PC COM 1-4
Agilent BenchLink Features
Configuration
Spreadsheet-like setup page
Upload and Download instrument setups
Computed channels using + - */, dB, dBm,
dBV, x
Graphical Displays
Real-time and historical data displays
Add, delete, size, and configure real time
Strip chart with markers and alarm indication,
bar and scatter charts, Histogram with
statistics, Bar meter, Digital meter, and
Data table
Graphical Controls
Sliders, switches, buttons, and LED lights
Alarm / Limit testing
Start/Stop scanning on alarm condition
Control 34903A relay state or 34907A
digital output on alarm
Data
Real time streamed (saved) to disk
Copy data or graphics to windows clipboard
Export user-selected data to .CVS, .XML,
or .TXT formats
Event logging
Automatic entry of alarms and errors
Enter user notes real time
Printing
Setup spreadsheet, all graphics, and
event log entries
Instrument Driver Support for Programming Languages
Universal
Compatible with Windows 95 and NT
Instrument Driver
[5]
Agilent VEE 3.2 or greater
Visual Basic 4.0,
LabWindows CVI 4.0,
LabVIEW 4.0
Labview Driver (VI)
LabVIEW 4.0
[1] Storage at temperatures above 40°C will decrease battery life
[2] Software provided on CD-ROM and includes utility to create floppy disks for
installation
[3] Interface and driver must be purchased and installed separately
[4] 90 MHz Pentium, 20 MB RAM
[5] Requires VISA command library for IEEE-488
Microsoft, Windows, and Windows NT are U.S. registered trademarks of
Microsoft Corporation.
Adobe and Acrobat are trademarks of Adobe Systems Incorporated.
Pentium is a U.S. registered trademark of Intel Corporation.
®
4.0 SP6a, 2000 SP4, XP
®
Acrobat
®
Reader V5.0 or higher
®
Internet Explorer V5.0 or higher
®
4, 800 MHz or
, √
1
2
x and full,
/
, or
bridge strain
/
1
2
4

Advertisement

Table of Contents
loading

Table of Contents