Series EPC-30 Current-to-Pressure (I/P)
Transducer
Introduction
Specifications
Item
Air supply input pressure
Air supply output pressure
Input impedance
Input signal
NOTE A: The transducer was calibrated using a 6.6−6.9 bar (95−100 psi) air supply.
B: With a 6.6−6.9 bar (95−100 psi) air supply, the output should be 0−0.3 bar (0−5 psi) with a 4 mA signal or
6.0−6.4 bar (87−93 psi) with a 20 mA signal.
E 2010 Nordson Corporation
WARNING! Allow only personnel with appropriate training and experience
to operate or service the equipment. The use of untrained or inexperienced
personnel to operate or service the equipment can result in injury, including
death, to themselves and others, and damage to the equipment.
This instruction sheet contains the procedure for installing the
current-to-pressure (I/P) transducer (kit P/N 772033).
If the Eclipset EPC-30 pattern control is equipped with an optional
input/output (I/O) board, the I/P transducer can be installed to provide
run-up control for the system. Refer to the Eclipse EPC-30 pattern control
manual for setting the run-up control.
The following table lists specifications for the I/P transducer.
Specification
8.6 − 10.3 bar (125 − 150 psi)
0.2 − 8.3 bar (1 − 120 psi)
260 ohms (nominal)
4 − 20 mA VDC
Instruction Sheet
P/N 331276B
Note
A
A
B
Part 331276B
02
02