Page 4
If any malfunction is discovered, please contact your nearest sales office or representative, or call TSI's Customer Service department at (800) 874-2811 (USA) or (001 651) 490-2811, International. Trademarks TSI, TSI logo, and PortaCount are registered trademarks of TSI Incorporated.
2. SETTING UP AND OPERATING THE MODEL 8026 PARTICLE GENERATOR ............3 Introduction ................3 Selecting a Fit Testing Location ..........3 Setting up the Model 8026 Particle Generator ......4 Operating the Model 8026 Particle Generator ......6 3. MAINTENANCE ................7 Routine Maintenance ..............7 Cleaning the Atomizer Nozzle ............
SA F ET Y IN F O R M A T IO N The Model 8026 Particle Generator includes a high-temperature safety switch. If the generator overheats, this switch shuts off the unit permanently, and it must be returned to TSI for servicing.
803011 Model 8026 Particle Generator Operation 1980319 and Service Manual Parts Identification for the Particle Generator Figure 1-1 identifies the parts of the Model 8026 Particle Generator. Please become familiar with these components before proceeding. Particle Generator Reservoir Figure 1-1...
Replacement Parts and Supplies The following items are sold by TSI as replacement parts and supplies for the Model 8026 Particle Generator. Table 1-2: Replacement Parts and Supplies Quantity Item Description Part/Model Package of 100 salt tablets 803011 Storage lid for reservoir...
This does not preclude using the Model 8026 Particle Generator for other applications to generate a moderate concentration of poly-disperse particles. Some general notes on alternate uses for the Model 8026 are found at the end of this chapter. Before setting up the Model 8026 Particle Generator, it is important to select a suitable location for fit testing.
Setting up the Model 8026 Particle Generator The particle generator supplements the naturally occurring particles in the room environment. Follow the steps below to set up the particle generator: 1. Fill the reservoir jar with clean tap water to the fill line. Use room temperature water to help the salt tablet dissolve more quickly.
Page 11
Never attempt to store the particle generator in the carrying case with a full or partially-filled reservoir attached, as the solution will leak out and damage the generator and/or carrying case. Setting Up and Operating the Model 8026 Particle Generator...
To increase particle concentration, adjust the output adjustment screw. Adjusting the Particle Generator Output The Model 8026 is set in the factory to a normal operating pressure. This will produce an aerosol output concentration that is appropriate for most environments. If higher counts are needed, increase the output by turning the output adjustment screw clockwise 1/2 turn.
Chapter 3 Maintenance Routine Maintenance The Model 8026 Particle Generator requires very little maintenance. Cleanliness is the single most important maintenance requirement. To ensure trouble-free operation of the generator, do the following: Use compressed air to blow out accumulated salt from inside the generator case.
Page 14
Figure 3-1 3. Use mild soap and warm water to clean the jet assembly parts. Examine the orifice for plugging. Rinse a second time if necessary. Either use clean compressed air to dry the parts, or air dry them. Note: The orifice is easily damaged. Do not insert anything in the orifice in an attempt to clean it.
Chapter 4 Troubleshooting This chapter describes the symptoms, possible causes and recommended solutions for common problems encountered with the Model 8026 Particle Generator. Some common problems that could be encountered include: Symptom Possible Cause Corrective Action Low particle The particle generator is Turn on the power to the not running.
Page 16
(This page intentionally left blank) Chapter 4...
Appendix A Specifications The Model 8026 Particle Generator specifications are as follows. Specifications are subject to change without notice. 10 in. x 4 in. x 5.25 in. Size ......... (25.4 cm x 10.1 cm x 13.3 cm) 2.75 lb (1.25 kg) Weight ........
Page 19
TSI Incorporated – Visit our website www.tsi.com for more information. Tel: +1 800 874 2811 India Tel: +91 80 67877200 Tel: +44 149 4 459200 China Tel: +86 10 8219 7688 France Tel: +33 1 41 19 21 99 Singapore Tel: +65 6595 6388...
Need help?
Do you have a question about the 8026 and is the answer not in the manual?