Operating Manual
Ergometer
custo ec5000
EXT 0001 - DK 1585
Version 001 – 2015-11-13
EL 20100043500

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Ergometer custo ec5000 and is the answer not in the manual?

Questions and answers

Summary of Contents for Custo Med Ergometer custo ec5000

  • Page 1 Operating Manual Ergometer custo ec5000 EXT 0001 - DK 1585 Version 001 – 2015-11-13 EL 20100043500...
  • Page 2 Manufacturer: ergoline GmbH Lindenstraße 5 72475 Bitz Germany Telefon: +49 (0) 7431 98 94 - 0 Fax: +49 (0) 7431 98 94 - 128 E-Mail: info@ergoline.com Internet: www.ergoline.com...
  • Page 3 Operating Manual Ergometer custo ec5000 © 2015 ergoline GmbH This Operating Manual may not be copied in its entirety or in part, duplicated in any form or by any means or translated into another language without the prior written consent of custo med GmbH and ergoline GmbH. The current version can be downloaded from this website: www.customed.de, in the section SUPPORT, Manuals.
  • Page 4 Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 5: Table Of Contents

    Specifi cations ........55 Electromagnetic Compatibility EN 60601-1-2 ....60 Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 6 Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 7: General Information

    Telefon: +49 (0) 7431 98 94 - 0 Fax: +49 (0) 7431 98 94 - 128 • On request custo med will provide a Field Service E-Mail: info@ergoline.com Manual. Internet: www.ergoline.com Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 8: Safety Information

    If in doubt, consult your local dealer or custo med. to electro magnetic interference. • IEC 60601-1+Al +A2:1995: 6.8.2.c, 19.2.b, 19.2.c, • IEC 60601-1:2005: 7.9.2.5, 8.1, 16.2.d, • MDD 93/42/EEC: Annex I clause 13.6.c Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 9: Intended Use

    General requirements measurement will also be aborted and the cuff defl ated. • EN 1060-3 Non-invasive sphygmomanometers, Part 3: Supplementary requirements for electro-mechanical blood pressure measuring systems Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 10: Symbols

    Ridlerstr. 65, 80339 München, Germany. Transport and storage label: Nationally Recognized Testing Laboratory do not stack. NRTL label for the USA and Canada. Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 11: Setup And Mains Connection

    Sockets for power cord and connection cables (under- side of ergometer) Power switch (toggle switch [ I / 0 ]) Saddle adjustment with clamping lever (optionally with gas pressure spring or motor) 5000 T OUCH Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 12: Transport

    In case of delicate fl ooring, it is recommended to place a mat under the ergometer to protect the fl ooring from damage by the feet. Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 13: Mounting The Control Terminal

    Before connecting the ergometer to the power line, check that the line voltage corresponds to the ratings on the type plate. The type plate is located on the back of the ergometer, at the bottom. Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 14 Removing the power cord results in a complete disconnection of the device from the power supply (all poles). Ensure that the power plug is readily accessible at all times. Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 15: Connecting The Ecg Cable

    Use only connection cables approved by custo med. A special PC driver software, which can be obtained from custo med, is required for operation via the USB port. Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 16: Connecting The Blood Pressure Cuff

    Within a few seconds the current oxygen saturation in percent (%) will be indicated on the display. O2 S Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 17: Preparing The Patient For Blood Pressure Measurements

    The cuff should be tight, but it should not constrict blood vessels. Make sure that the cuff cannot shift when the patient moves during the exercise test. ICROPHONE PLACEMENT ON THE ARTERY Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 18: Checking The Cuff Tubing

    Therefore, the computer aborts the measurement, if a minimum The same measures are recommended, if the cuff does not pressure is not attained within a few seconds. defl ate correctly. Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 19: Operation

    • If, during an exercise test, the speed drops below 30 RPM, the load readout starts blinking on the display and the load is reduced to zero. Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 20: Handlebar Adjustment

    Press the [ Saddle ] or the key to enable the adjust- ment of the saddle: the adjustment keys will be displayed (or via the buttons on the control terminal). Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 21: Control Terminal Smart

    PC MODE An external device (e.g., an ECG recorder, a PC-based ECG system) controls the ergometer - no intervention at all is required at the ergometer. Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 22: Pc Mode

    BP measurement (BPM) pedal speed (RPM) Note • During an exercise test, the saddle height can be adjusted with mmHg ♥  /   m in Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 23: Control Terminal Plus

    The functions of these three softkeys chan ge with the displayed menu - the key label describing the EYPAD function is shown on the display. Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 24: Operating Modes With Control Terminal Plus

    (see chapter "Settings") MANUAL The ergometer is controlled manually, i.e., the user performs all load changes via the keypad. SETTINGS Used to configure the ergometer. Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 25: Pc Mode

    To reactivate the saddle height adjustment function, press and the arrow keys will again be displayed. • Additional blood pressure measurements can be initiated with mmHg ♥  /   m in Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 26: Ergometry

    (in increments of +/-1 W up to +/-25 W, as con- figured). mmHg ♥  /   m in + 5 W Recovery - 5 W CREEN DISPLAY DURING THE TEST Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 27 It is recommended that the patient continue to pedal in the recovery phase. + 5 W - 5 W The END key in the middle will terminate the test. ECOVERY PHASE Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 28: Manual

    There is no recovery phase in the manual mode. mmHg ♥  /   m in + 5 W - 5 W CREEN DISPLAY DURING THE TEST Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 29: Settings With Control Terminal Plus

    Use the softkeys on the right and left (↑ ↓) to position the bar cursor on the protocol to change (No. 6 - 15) and confirm the selection with SELECT. ELECTING THE EXERCISE TEST PROTOCOL TO EDIT Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 30 +/- 1, 5, 10 und 25 Watts. + / - 5 Watt + / - 10 Watt + / - 25 Watt Select ELECTING THE INCREMENT FOR MANUAL LOAD CHANGES Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 31 ETTING THE DATE Date 15  .   1 2  .   2 014 Time 09  :   0 0  :   0 0 Select ETTING THE DAY Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 32 • The “Analog/Digital” and “Digital” communication is only possible when PC Mode is selected from the main menu or when this is the default mode. Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 33 > 125 951 - 999 > 130 ULSE ISPLAY The pulse readout on the display can be turned off. Display Puls Display Select ETTING THE PULSE READOUT Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 34 Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 35: Control Terminal Touch

    The ergometer runs a self-test. Subsequently, the main menu displays. custo med GmbH ec5000 TEST SCREEN The ergometer software is controlled from the touch panel. PC Mode Ergometry Training/Test Manual Setup Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 36: Operating Modes With Control Terminal Touch

    MANUAL The ergometer is controlled manually, i.e., the user per- Manual forms all load changes via the touch panel. SETUP Used to configure the ergometer. Setup Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 37: Pc Mode

    MODE SCREEN A blood pressure measurement can be initiated with the [ RR ] key. Pressing the [ RR ] key a second time during a measurement will stop the measurement. Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 38: Ergometry

    [ Setup ] key in the main menu. Description Programm 6 Type Step Basic Load 30 W Stage Time 2 min 25 W Stage Rate XERCISE TEST PROTOCOL Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 39 To change the load, press the [ + 5 W ] and [ - 5 W ] keys (the increments can be configured from +/-1 W up to +/-25 W as needed). TARTING AN EXERCISE TEST Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 40 When you touch [ Stop ] again, the training will be termi- nated. - 5 W --/-- 00:00 1/min mmHG mm:ss 1/min Stop 1/min const. + 5 W - 5 W ERMINATING AN EXERCISE TEST Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 41: Manual

    [ Stop ] key. 1/min The load will immediately drop to 0 watt. There is no recovery phase in the manual mode. ERMINATING A MANUAL TEST Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 42: Setup With Control Terminal

    · PC Mode · Ergometry Menu · Training/Test PC Mode · Manual Ergometry Confirm the selection with the [   ] key. Training/Test Manual EFAULT MODE SETUP Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 43 (from 6 to 100 W) and the load Stage Time 2 min increase (from 1 to 50 W). 25 W Stage Rate Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 44 With the [   2. ] and [   1. ] keys, you toggle between the 25 W Stage Rate different screens. Recovery Load 25 W Recovery Time 2 min Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 45 The entire communication (load control and BP measure- ments) is controlled with analog signals. No digital data will be sent. Choose the appropriate communication mode and confirm with the [   ] key. Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 46 Recovery Time 3 min Inputs are confirmed with the [   ] key. Load Change Stage Time 20 sec Weight 3 Kg Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 47 10: ( ID=--- ) 14: ( ID=--- ) 11: ( ID=--- ) 15: ( ID=--- ) 12: ( ID=--- ) 16: ( ID=--- ) HR B UMBER SETUP Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 48 Duration: load + D uration: load + 3 min 3 min Duration: load - D uration: load - 3 min 3 min ONTROL SETUP Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 49 This menu shows the software version and the date of the technical inspection of the measuring system (MTK). Software Version EF 10 MTK: 15.05.2015 OFTWARE ERSION SETUP Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 50 Switch the pulse readout on or off. Display Select the blood pressure unit: mmHg (millimeter of mercury) or kPa (kilopascal). Pulse Display NIBP mmHg / kPa mmHg ETUP SCREEN Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 51: Cleaning, Maintenance, Disposal

    Do not use acids, alkaline solutions (household cleaners) or caustic disinfectants. LEANING THE ADDLE Clean the saddle with a soft and dry or moist cloth. Disin- fectants used should not contain any alcohol. Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 52 Velcro tab. Then close the tab. Fold the end of the cuff over and introduce it into the metal clasp. NSERTING THE ICROPHONE Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 53 Please contact an authorized representative of the manu- facturer for information concerning the decommissioning of your equipment. There is no waste approval. Proper disposal is documented by custo med. Consult Operating Manual! Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 54 Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 55: Specifications

    120 cm and 210 cm continuous handlebar adjustment over 350° Handlebar height adjustment version 1: rigid steering column (standard) version 2: electrical adjustment of the steering column (option) Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 56 800 to 1060 hPa transport and storage: temperature: -20 to +70 °C rel. humidity 10 to 95%, no condensation atmospheric pressure: 500 to 1060 hPa Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 57 300 mmHg Cuff deflation method pulse-dependent deflation rate approx. 3 mmHg/beat or approx. 3 mmHg/s Calibration calibration with external pressure meter Artifact rejection automatic artifact rejection Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 58 [sec] [min] ramping pro- tocol PWC-130 PWC-150 PWC-170 (*) the protocol advances to the recovery phase as soon as the target heart rate (130/150/170) is reached Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 59 IEC 60601-1 AMILY OF CHARACTERISTICS OF THE LOAD PERIODS ACCORDING TO Under permanent load, the load periods and pauses (white) shall be observed. Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 60: Electromagnetic Compatibility En 60601-1-2

    Harmonic emissions to EN 61000-3-2 Class A nected to the public low-voltage power supply network that supplies buildings used for domestic purposes. Voltage fl uctuations/fl icker emissions Complies to EN 61000-3-3 Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 61 The ec5000 ergometer has no components suscep- tible to magnetic fi elds. NOTE: UT is the a.c. mains voltage prior to application of the test level. Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 62 Over the frequency range from 150 kHz to 80 MHz, fi eld strengths should be less than 3 V/m. Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 63 At 80 MHz and 800 MHz, the higher frequency range applies NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is aff ected by absorption and refl ection from structures, objects, and people. Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 64 Saugleitungen 705.375 Analog-Schnittstelle (COM-Modul) 705.362 Puls-Brustgurt ERGOLINE 705.295 Puls-Brustgurt POLAR 705.084 Sattelverstellung horizontal 471.107 Rennsattel mit Standardaufnahme 705.085 Pedal-Kurbeln, verstellbar is in conformity with the Directive Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 65 Class Classification according to: Appendix IX / rule 10 Beginn der Gültigkeit / Begin of the validity: /Date: 03.07.2015 /City: Bitz den, 03.07.2015 Axel Bodmer Ergometer custo ec5000 | EXT 0001 – DK 1585 | Version 001 – 2015-11-13...
  • Page 66 Manufacturer: Distribution: ergoline GmbH custo med GmbH Lindenstraße 5 Leibnizstraße 7 72475 Bitz 85521 Ottobrunn Germany Germany Telefon: +49 (0) 7431 98 94 - 0 Phone: +49 (0) 89 710 98 - 00 Fax: +49 (0) 7431 98 94 - 128 Fax: +49 (0) 89 710 98 - 10 E-Mail:...