Supply Voltage - TE LDM-1000 Operation Manual

Lvdt/rvdt signal conditioning module
Table of Contents

Advertisement

LDM-1000
LVDT/RVDT Signal Conditioning Module
The list below is the minimum information required to perform a successful calibration:
Recommended LVDT or RVDT excitation frequency (to set the oscillator frequency)
LVDT or RVDT Output sensitivity at the excitation frequency (to set the gain)
Primary (input) impedance at the excitation frequency (to set the oscillator output voltage)
The ± full scale linear or angular range over which you intend to perform the calibration
Analog output signal required by your application

4. 1. Supply Voltage

The LDM-1000 supply voltage range is 18~30 VDC with the JP1 jumper in the storage position (as shipped position). Only this
range (18~30VDC) can be used for 4-20mA output operation. The operating voltage may be changed to 10~18 VDC by installing
internal jumper JP1 across both pins. See the diagrams below for jumper settings.
CAUTION: Operating the unit at voltages above +18 volts with the jumper in the low voltage position will overheat
the internal voltage regulators and reduce the operational life of the LDM-1000.
TE CONNECTIVITY SENSORS
/// LDM-1000 OPERATION MANUAL
P/N 09290100-000 REV. C
05/2016
Page 5