Oxford Stature User Instruction Manual page 16

Patient lift
Table of Contents

Advertisement

Available languages

Available languages

Oxford
®
Stature
6. Mechanical Emergency Down
In the case of a complete electrical failure the Stature lift is fitted with mechanical lowering device.
This device is located on the left hand side of the lift's chassis.
The lowering mechanism is operated by inserting a 6mm Hex socket into the aperture and onto
the mechanism turning clockwise to raise the lift and anticlockwise to lower the lift. This can be
done manually with a socket wrench.
RECOMMENDATION:
In order to raise or lower the boom more quickly it is recommended that a 6mm Hex socket be
used in conjunction with a 1/4 inch drive adaptor and an electric reversible drill (NOTE: do not
use hammer setting). This will increase the speed in which the patient can be safely raised and
lowered. A 6mm Hex socket is contained in the top cap located on the top of the mast.
NOTE: Before raising or lowering the patient, ensure you have you have planned your
destination carefully. For instance back onto a chair or bed.
7. Batteries
The batteries are protected from deep discharge by a LOW VOLTAGE ALARM. This will sound
when the batteries need recharging and the hand control is being operated. It will not sound
independently of the hand control being operated. DO NOT IGNORE THIS WARNING ALARM.
Complete the lifting operation and place the battery on charge (see charging instructions).
16