new brunswick innova 2000 Operating Manual page 24

Stackable incubator shaker with optional refrigeration
Table of Contents

Advertisement

Operation
24
New Brunswick Innova
English (EN)
Abb. 5-1: Recorder connector as seen from the rear of the unit
5
4
3
8
7
9
Fig. 5-1:
Recorder connector as seen from the
rear of the unit
Pin number
6
2
7
3
®
2000/2050 Shaker
2
1
6
Signal name
Speed
Ground
Temperature
Ground
• For speed, each channel should have signal
conditioning that accepts 0 – 5 V input.
• For temperature, each chanel should have a
signal conditioning that accepts 0 – 5 V input
• A mating connector is required on the recorder
cable (not supplied, but available from
Eppendorf). This is a 9-pin male D subminiature
connector, AMP Amplimite HDP-20 series or
equivalent.
Scale
1 V = 100 rpm
1 V = 20 °C