Certification ....................Technical description....................Design and function ..................System components..................4.2.1 Components of the inEos Blue............4.2.2 Accessories (included in the scope of supply) ........4.2.3 Calibration parts (included in scope of supply) ........4.2.3.1 3D acquisition system ..............4.2.3.2 XYZ axis alignment ................
Page 3
Installing the software................... Connecting the unit to the PC ..............Switching the unit on and starting the software..........Adding inEos Blue to the device configuration ..........Connecting the unit....................Connecting the inEos Blue to the PC ............Connecting the inEos Blue to the power supply...........
Page 4
Sirona Dental Systems GmbH Operating Instructions inEos Blue 11.7.1.4Buccal registration................11.7.1.5Terminating acquisition ..............11.7.2 Acquisition of a crown cap ..............Cleaning and care ....................12.1 Care of the 3D camera system ..............12.2 Cleaning the outer surface ................12.3 Protection against medicaments ..............
Operating Instructions inEos Blue Dear Customer, Dear Customer, ® Thank you for purchasing your inEos Blue from Sirona. Improper use and handling can create hazards and cause damage. Please read and follow these operating instructions carefully and always keep them within easy reach.
2 General information Sirona Dental Systems GmbH 2.1 Structure of the document Operating Instructions inEos Blue General information General information Please read this document completely and follow the instructions exactly. You should always keep it within reach. Original language of the present document: German.
Sirona Dental Systems GmbH 2 General information Operating Instructions inEos Blue 2.2 Legend 2.1.2 Formats and symbols used Formats and symbols used The formats and symbols used in this document have the following meaning: ✔ Prompts you to do something.
Warning: inEos connected to CEREC 3 acquisition unit WARNING No acquisitions in the patient's mouth As long as an inEos Blue is connected to a CEREC 3 acquisition unit, no acquisitions may be made in the patient's mouth using the connected CEREC 3D camera.
WARNING No acquisitions in the patient's mouth within the patient environment As long as an inEos Blue is connected to a CEREC AC acquisition unit, no acquisitions may be made in the patient's mouth using the connected CEREC Bluecam if the inEos Blue is located outside of the patient environment.
3.1.3 Connection of external equipment Connection of external equipment If any equipment not approved by Sirona is connected, it must comply with the applicable standards: ● EN 60 950 for information technology equipment, and ● EN 61 010-1 for laboratory equipment.
Operating Instructions inEos Blue 3.2 Intended use Intended use Intended use Intended use of the inEos Blue This system is used for 3D digitization of single tooth or complete jaw saw-cut models and impressions. Follow the operating instructions If the unit is used for any usage purpose other than the one mentioned above, it may be damaged.
3 Safety Sirona Dental Systems GmbH 3.3 Certification Operating Instructions inEos Blue Certification Certification CE mark This product bears the CE mark in accordance with the provisions of directives 2006/95/EC (Low Voltage Directive) and 2004/108/EC (EMC Directive). CAUTION CE mark for connected products Further products which are connected to this unit must also bear the CE mark.
Design and function Design and function The inEos Blue is designed as a desktop unit and consists of several components (see Chapter "System components" [ → 13]). It is powered from the standard local power network via an external power supply unit.
4 Technical description Sirona Dental Systems GmbH 4.2 System components Operating Instructions inEos Blue 4.2.1 Components of the inEos Blue Components of the inEos Blue Vertical unit and front side of base plate Vertical unit Base plate Rotary knobs for height...
Sirona Dental Systems GmbH 4 Technical description Operating Instructions inEos Blue 4.2 System components Rear side of base plate Main switch Supply voltage input USB port Rotation mouse 4.2.2 Accessories (included in the scope of supply) Accessories (included in the scope of supply)
4 Technical description Sirona Dental Systems GmbH 4.2 System components Operating Instructions inEos Blue Foot control 4.2.3 Calibration parts (included in scope of supply) Calibration parts (included in scope of supply) 3D acquisition system 4.2.3.1 3D acquisition system Calibration part for calibrating the 3D acquisition system.
Sirona Dental Systems GmbH 4 Technical description Operating Instructions inEos Blue 4.3 Technical data Technical data Technical data Model designation inEos Blue Power connection via power supply unit: INPUT: 100-240VAC / 47- 63Hz / 1.0 A max. permissible line voltage fluctuations: ±...
Install inEos Blue in a protected area 3D image acquisition may be adversely affected by bright light. Set the inEos Blue up so that it is not located directly in the beam path of an extreme light source and not exposed to direct sunlight.
The software is installed. Connecting the unit to the PC Connecting the unit to the PC ➢ Connect the inEos Blue to the PC (see also section "Connecting the unit" [ → 21]). 62 99 205 D3446 D3446.201.02.04.02 03.2012...
If no inEos Blue is displayed, check the configuration of the interface. 5. Enter a name for the new unit. IMPORTANT Calibrating the system After installing the inEos Blue, you must calibrate the system (see the chapter entitled "Calibration" [ → 27]). 62 99 205 D3446 D3446.201.02.04.02 03.2012...
"inEos Blue" Socket (USB port) on the rear of the PC (labeled ➢ Connect the socket (USB port) of the inEos Blue (A) to the socket (USB port) of your PC (C) via the interface cable included in the scope of supply (B).
Connecting the inEos Blue to the power supply Connecting the inEos Blue to the power supply Socket for power supply ➢ Connect the socket (A) of the inEos Blue to the power supply via the power supply unit included in the scope of supply. NOTICE Use a grounded power outlet The power supply unit must be connected to a grounded power outlet.
Operating Instructions inEos Blue 7.3 Connecting the rotation mouse Connecting the rotation mouse Connecting the rotation mouse Socket of the inEos Blue Rotation mouse ➢ Plug the connector of the rotation mouse (B) into the matching socket (A) on the inEos Blue.
Controls and displays Controls and displays Controls on the inEos Blue Controls on the inEos Blue The following controls are available on the inEos Blue: Rotary knobs for height adjustment (focusing) of the lens on the vertical unit Start button...
Sirona Dental Systems GmbH 8 Controls and displays Operating Instructions inEos Blue 8.2 Ready LED Ready LED Ready LED The blue Ready LED indicates the operating status of the unit: LED not lit: Unit is switched off LED flashes: Unit is switched on and ready for operation 62 99 205 D3446 D3446.201.02.04.02 03.2012...
Switching the system on Switching the inEos Blue on 1. Switch on the PC. 2. Switch the inEos Blue on at the main switch. Starting the software ➢ Start the software at the PC via the Windows start menu. 62 99 205 D3446...
● X-Y-Z axis alignment NOTICE Calibrate prior to initial startup Prior to initial startup of the inEos Blue, you must perform a full calibration. No further calibrations are usually required afterwards. If you remove the shifting plate, the X-Y-Z alignment must be recalibrated after reassembling the unit.
10.3 Calibrating the XYZ axis alignment Calibrating the XYZ axis alignment Preparing the inEos Blue ✔ The PC and the inEos Blue are switched on. ✔ The inLab SW software has been started 1. Open the system menu. "Configuration" 2. Click on the button.
Page 29
Sirona Dental Systems GmbH 10 Calibration Operating Instructions inEos Blue 10.3 Calibrating the XYZ axis alignment NOTICE Do not change this position The position of the vertical unit must not be further changed until step 5. "OK" 3. Click the button.
Page 30
10 Calibration Sirona Dental Systems GmbH 10.3 Calibrating the XYZ axis alignment Operating Instructions inEos Blue Step 4 Step 4 1. Place the calibration plate in position 4. Note the position of the beveled corner. The grid balls must be seated as shown in the illustration.
Page 31
Sirona Dental Systems GmbH 10 Calibration Operating Instructions inEos Blue 10.3 Calibrating the XYZ axis alignment Step 6 Step 6 1. Turn the calibration plate counterclockwise 90° and place it in position 3. Note the position of the beveled corner. The grid balls must be seated as shown in the illustration.
The following are suitable for this purpose: CEREC Optispray (REF 61 44 179) For scanning with inEos Blue, we recommend using CEREC Stone BC plaster (REF 62 37 502). Saw-cut etc. ● When preparing saw-cut models, be careful not to mill off the single stump segments.
Page 33
Operating the foot control Operating the foot control You can use the foot switch instead of the start button of the inEos Blue. Link to inEos Blue To do this, connect the foot switch as described in the chapter entitled "Connecting the foot switch"...
Switching the inEos Blue on/off The inEos Blue is turned on automatically when you switch to the acquisition phase. If the inEos Blue is not used for some time, it will switch off automatically. ➢ Click on the inEos icon to switch the inEos Blue on or off.
Page 35
Focusing the image. Do not touch the rotary buttons while the scans are being taken. 5. Single-click the Start button of the inEos Blue. The image is acquired. 6. Add additional scans from adjacent areas: Move the model in the desired direction by one stop position and repeat steps 4 to 5.
If you want to interrupt the automatic scan, single-click the Start button of the inEos Blue. The crosshairs turn green. Use the Start button to acquire additional exposures.
Page 37
3. Focus the image with the rotary knobs for height adjustment; see Focusing the image. 4. Single-click the Start button of the inEos Blue. The first scan is taken and the crosshairs are displayed blue. 5. Add additional scans from adjacent areas: Move the model freely on the table until you have reached the next acquisition position.
Tip: You can turn the entire model on the shifting plate by 180° if necessary. 2. Double-click the Start button of the inEos Blue. An image is acquired. 3. Repeat steps 1 and 2 until all of the required scans have been taken.
2. Tighten the fastening screw until the model is fixed in place. 3. Align the platform in such a way that the model is horizontal. 4. Place the model holder on the shifting plate of the inEos Blue. The oral side points to the column of the inEos Blue.
Page 40
4. Focus the image with the rotary knobs for height adjustment; see Focusing the image. 5. Single-click the Start button of the inEos Blue to start the automatic scan. The first scan is taken and the crosshairs are displayed blue. If you shift the model or re-focus it, additional scans will automatically be generated.
Tip: In the preview, you can rotate the model. In this way, you can determine whether the information is complete. 2. Click the Start button of the inEos Blue twice to create single scans. The overview image is supplemented.
5. Focus the image with the rotary knobs for height adjustment; see Focusing the image. 6. Single-click the Start button of the inEos Blue to start the automatic scan. The first scan is taken and the crosshairs are displayed blue. If you shift the model or re-focus it, additional scans will automatically be generated.
3. Rotate the rotation pot until the object remains in the image field in every position. 4. Focus the image with the rotary knobs for height adjustment. 5. Single-click the Start button of the inEos Blue to start the scan. Eight scans will be acquired. Terminating acquisition Terminating acquisition 1.
12 Cleaning and care Sirona Dental Systems GmbH 12.1 Care of the 3D camera system Operating Instructions inEos Blue Cleaning and care Cleaning and care 12.1 Care of the 3D camera system Care of the 3D camera system The 3D camera system is a very sensitive optical device and must therefore be handled with the utmost care.