34924A 70-Channel Reed Multiplexer

34924T Terminal Blocks

Two terminal blocks are available to facilitate wiring connections to the
34924A module:
34924T-001 (Option 001)
34924T-002 (Option 002)
The 34980A Product Reference CD (shipped with the instrument) contains
a 34924T Wiring Log for you to document your wiring configuration for
this module. You can open the wiring log file in Microsoft® Excel® or
Adobe® Acrobat® format.
34924T-001 Terminal Block
This terminal block with solder- type connections is labeled with the model
number and the abbreviated module name. In addition, space is available
on the label for you to write the slot number.
32
Terminal block with solder connectors
Terminal block with screw connectors
All modules that connect to the internal DMM are interlock
N O T E
protected. This means that when an installed module is exposed
(no terminal block or cable is connected), the Analog Bus relays are
open and disconnected from the Analog Buses. See
Interlock"
Agilent 34921A-34925A Low Frequency Multiplexer Modules User's Guide
on page 3 for further information.