HYDROGUARD
HG-702
®
TurbiPlus
Water Quality Analyzer &
Controller
User Manual
Document ID HG702TPUM1.3

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the HYDROGUARD HG-702 TURBIPLUS and is the answer not in the manual?

Questions and answers

Summary of Contents for Blue I HYDROGUARD HG-702 TURBIPLUS

  • Page 1 ® HYDROGUARD HG-702 ® TurbiPlus Water Quality Analyzer & Controller User Manual Document ID HG702TPUM1.3...
  • Page 2 TurbiPlus are registered trademarks of Blue I Water Technologies Ltd. Copyright © 2014 by Blue I Water Technologies Ltd. Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 3: Table Of Contents

    Cleaning the Filter ....................40 Shut-Down and Winterizing ..................42 Start-up and Preventive Maintenance ..............42 6.5.1 Replacing Pump Head and Tubes ............... 43 Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 4 Setting Proportional Factor ..................72 Setting Pump Period ....................74 Step By Step Proportional Settings ................. 74 Technical Specifications ....................75 Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 5 Table 10 - Troubleshooting: Problems and Suggested Solutions ........... 46 Table 11 - Additional features in Operator Menu (With Total Chlorine Monitoring) ....56 Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com...
  • Page 6 Table 15 - Modbus Configuration Options ................65 Table 16 - Modbus Communications Options ................. 66 Table 17 - Relays and their controlled dosing systems ............72 Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 7: General Safety Precautions

    220-240VAC requiert un fusible de 0,5 Amp. completed prior to wiring. Ces modifications doivent être accomplies avant le branchement électrique. Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 8 être effectuée que par un technicien ayant reçu la formation adéquate et dûment habilité. Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 9: Preface

    This manual is for qualified and trained service technicians who will install and service the HYDROGUARD HG-702 TurbiPlus Water Quality Analyzer. It provides instructions on how to install the HYDROGUARD system, how to integrate it with external chemical dosing systems and how to calibrate, operate and maintain the system.
  • Page 10: Overview

    Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com...
  • Page 11: Measurements And Features

    3.2 Measurements and Features The HYDROGUARD HG-702 TurbiPlus can be configured to measure any combination of the following water quality parameters:  Free chlorine  Total chlorine  Both free chlorine and total chlorine  Turbidity Optional measurements:  Temperature ...
  • Page 12 4-20 flow meter. Internal 4-20 Output Module* – Provides up to six 4-20 mA outputs for any measured variable. *Optional module Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 13: Figure 1: Hydroguard Hg-702 Turbiplus Components

    Figure 1: HYDROGUARD HG-702 TurbiPlus Components Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 14 Probe ORP Probe Temp Probe (PT-100) Rotating Flow Switch LED Module Pressurized Sampling Cell Colorimeter Drain Colorimeter Drain Funnel Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 15: Installation

    Note: During disassembly, be sure to retain all hardware for use during reassembly. Upgrade and replacement kits may re-use some original components. Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com...
  • Page 16: Mechanical Installation Requirements

    Minimum inlet pressure should be 4.4 psi (0.3 bar) and should not exceed 14.5 psi (1 bar) using a pressure reducer attached to the pre-filter outlet (Figure 3). Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com...
  • Page 17: Colorimeter Drainage

    If the colorimeter drains to a bucket or basin, the end of the drainage pipe should terminate above the bucket or basin rim. Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com...
  • Page 18: Electrical Requirements And Installation

    2. Connect the line (live) wire to the electronic relay card connector marked Line. 3. Connect the neutral wire to the electronic relay card connector marked Neutral. Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com...
  • Page 19: Figure 6: 110-120Vac Set Up

    For information on wiring and using the relays as dry contact or for control, see section 7.8 below: Chlorine Shock Mode. Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com...
  • Page 20: Input Switches

    5 First Time Operation and Calibration 5.1 Installing Additional Sensors Install all additional sensors and meters and connect to the main system of the HYDROGUARD HG-702 TurbiPlus, following the instructions for each sensor as follows. 5.1.1 pH, ORP (Redox) and Temperature Measurements 5.1.1.1 Installation If the HYDROGUARD system was ordered with these components some steps in this process will have been completed in the factory and may be skipped.
  • Page 21 If either one or both indicators do not display a proper reading or are not stable, confirm proper installation and then see the Troubleshooting section (6.6). Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com...
  • Page 22: Installing Reagents

    6. Repeat five (5) times until all the indicator salt is dissolved in the indicator fluid. See illustration below. Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 23: Figure 9: Mixing Indicator Salt Into Indicator Fluid

    Do not refill the reagent bottles. Do not mix or add reagent from other bottles. Only use original Blue I Water Technologies reagents. Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 24: First Time Menu Setup

    The HYDROGUARD control panel, Figure 12, is a simple, intuitive interface for monitoring and controlling water quality with the following components: Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com...
  • Page 25: Menus

    The Technician Menu includes settings and calibrations that should be restricted to specially trained HYDROGUARD maintenance technicians. Each menu has a separate password. Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 26: Operator Menu

    Initiates an alarm when Total Cl is above value Comb. Cl high alarm* Initiates an alarm when combined (Free & total ) Cl is above value Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 27: Table 3 - Operator Menu Functions - Minimum And Maximum Values

    Turbidity set point 0.60 NTUI calibrated to sensor value was NTUh calibrated to sensor value was Turbidity high alarm 99.9 99.9 Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 28: Configuring Settings In The Operator Menu

    The second row of the LCD display, below the value that is being changed, shows the current value. Press Enter to save the new setting or Esc to abort without saving the new setting. Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 29: Technician Menu

    Cl is being maintained at low levels. Therefore, this alarm feature may be turned off. For cases when the Cl level is always expected to be above 0.5ppm, Blue I Water Technologies highly recommends that this alarm remain in use except for start-up.
  • Page 30: Table 5 - Technician Menu And Variable Limits

    View free chlor View pH* View ORP* View conductivity* Address Ver. Max flow range 200 / 999 /hour or GPM Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 31: Configuring Settings In The Technician Menu

    It is recommended that calibration be performed every 6 to 12 months. Minor deviations will exist between all testing equipment. These minor differences do not warrant the calibration of the colorimeter. Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 32: Chlorine Calibration

    The “Calibrate Cl to” value is the new value that you want to set. Figure 13: Sensor Reading After Calibration Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 33: Ph Calibration

    24 hours. Note: The temperature probe provides grounding and stabilizes the pH and ORP readings. Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 34: Orp (Redox) Calibration

    8. Press OK again. 9. Press the up or down arrow until the value is the same as the independent digital thermometer. Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 35: Turbidity Calibration

    If pH and ORP are reading slightly low, do not calibrate as they will continue to increase for the first 24-48 hours. Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com...
  • Page 36: Routine Operation And Maintenance

    This allows the two 600 ml bottles last approximately 4 to 8 weeks. A Low Reagents alarm is issued when Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com...
  • Page 37: Monitoring Hydroguard Alarms

    The water flow rate in the main Stop all chemical Low flow circulation pipe is lower than the flow dosing (all relays are limit. open). Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 38: Replacing Reagents

    The blue LED lights in the sampling cell will flash continuously until the reagents are replaced or fully depleted. When depleted a NO REAGENTS alarm is issued and the chlorine Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com...
  • Page 39: Figure 14: Press Tab Above Siphon Cover And Pull Bottle Down

    Replace the reagent injector into its position 7. Close the analyzing unit door. 8. Locate the LCD detail readings window in the HYDROGUARD control panel. Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 40: Cleaning The Filter

    A. Push the air relief button (1) for the filter top to reduce the vacuum and unscrew the filter casing (2). B. Remove the cartridge from the filter casing (3). Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 41: Figure 16: Pull Filter From Casing

    B. Verify the casing’s O-ring is in place and screw casing to filter base, until the gasket is firmly compressed against the filter connector pipe. 7. Open the water supply valve. Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 42: Shut-Down And Winterizing

    (See Section 5.2.8) 5. If reagents are not dripping as they should, replace the colorimeter cap with a new one. Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com...
  • Page 43: Replacing Pump Head And Tubes

    Table 10 outlines some possible problems that may occur but will not generate an alarm, their potential causes and solutions. Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 44: Table 9 - Troubleshooting: Alarm Descriptions And Solutions

    ORP above upper limit. safety against over- poor grounding. chlorination. Increase high/low ORP level if operation is normal. Replace probe if needed. Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 45 Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 46: Table 10 - Troubleshooting: Problems And Suggested Solutions

    * No dosing only affects the relay operation. Alarm relay will close and all other relays will open. Table 10 - Troubleshooting: Problems and Suggested Solutions Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com...
  • Page 47: Replacing Components

    I Water Technologies Ltd. or the qualified representative who supplied the product may void the warranty. Blue I Water Technologies Ltd. takes no responsibility, written or implied, for installation or maintenance of HYDROGUARD that is not performed by a properly trained and certified HYDROGUARD technician.
  • Page 48: Replacing The Temperature Probe

    4. Install the new reagent siphons: a. Attach the four (4) mounting screws. b. Connect the two (2) red wires to the colorimeter card. Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 49: Replacing The Reagent Level Switch

    Locate the colorimeter assembly unit in the analyzing module (Part number 12 in Figure 1). Figure 18: Colorimeter Assembly Unit Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 50: Replacing The Solenoid Valve

    9. Complete the process by reattaching the pressurized sample cell in place with the four (4) screws, and reconnecting the solenoid valve cable to the colorimeter card. Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com...
  • Page 51: Replacing The Control Panel Module (Electronics Card)

    3. Put in the new card and tighten the 4 mounting screws. 4. Connect the flat cable plug to the card. Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com...
  • Page 52: Replacing The I/O Module

    4. Connect the flat cable plug to the colorimeter card. 5. If removed, replace and reconnect the pH, Redox and Temperature sensors card as described in section 6.7.11. Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 53: Optional Features

    Chip for control panel card with software version FAT (Free And Total)  DPD3 reagent A diagram of a Free and Total Chlorine analyzer is presented in Figure 21. Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 54: Figure 21: Free And Total Chlorine Controller

    Figure 21: Free and Total Chlorine Controller Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 55 2. Remove the white rotor cover and release the tube. 3. Feed the new tube around the metal pump pins. Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 56: Additional Menus And Settings

    Table 11 - Additional features in Operator Menu (With Total Chlorine Monitoring) Table 12 - Additional Alarms (With Total Chlorine Monitoring) Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com...
  • Page 57: Table 13 - Alarms (With Total Chlorine Monitoring)

    High NTU Turbidity above high limit EXTERNAL OFF External flow switch is off Total Cl high Total Cl above high limit Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 58: Table 14 - Technician Menu (With Total Chlorine Monitoring)

    Will turn display of Flow rate on/off (only use if flow meter connected) Flow Rate Choose between metric and US units Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 59: Total Chlorine Monitoring Setup

    4. Press OK 5. Insert password (default 456) and press OK 6. Press up or down button to change function to “On” Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 60: Ph, Orp And Temperature Measurements

    1) Attach the conductivity card to the inside of the control panel door below the control panel module using the four (4) supplied screws. Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com...
  • Page 61: First Time Set-Up And General Operation

    7.4.3.3 Replacing the Conductivity Sensor 1. Turn off the inlet and outlet water to the sampling cell and the power to the analyzer. Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com...
  • Page 62: Shut-Down And Winterizing

    7.5 Flow Meter Blue I Water Technologies does not supply flow meters. However, the analyzer supports a flow meter connection with input connections and display. 7.5.1 Installation 7.5.1.1 Required Components...
  • Page 63: Routine Flow Meter Maintenance And Troubleshooting

    7.5.2 Routine Flow Meter Maintenance and Troubleshooting Follow manufacture’s recommendations for maintenance procedures and any troubleshooting issues for a flow meter. Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 64: Modbus Communication Protocol

    The second row of the menu display, below the value that is being changed, shows the current value. Press Enter to save the new setting or Esc to abort without saving the new setting. Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 65: Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel

    2 stop / Parity / 19200bps / Parity Even / value / Swapped 9600bps 1 stop bit No parity floating point Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 66: Table 16 - Modbus Communications Options

    Conductivity 4-20(1) Discrete Input 10027 On/Off No Flow Discrete Input 10032 Low Flow Discrete Input 10033 No Reagents Discrete Input 10034 Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 67 - Replace light bit7 - Colorimeter communication Error bit8 - Piston stuck Controller ID Input Register 30019 16 bit serial number Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 68: Communication Options

    20 mA = Max Cl dosing  Inverted: 20 mA = No Cl dosing 4 mA = Max Cl dosing Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 69 (an output is inactive if it does not have a chipset installed next to the terminal block). 2) Record the current in mA. 3) Confirm that the value is accurate using the following equation. Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 70: Chlorine Shock Mode

    This is only required if the relays are being used to power and control external dosing equipment. The relays will act as dry contacts if no power is supplied. A dependent power Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com...
  • Page 71: Wiring To Dosing Systems

    4 amps. Figure 22: Relay positions on board The following table lists the relays and the dosing systems they control: Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 72: Setting Proportional Control

    In large systems, changes in chemical levels occur slowly. The dosing systems must feed large amounts of chemicals for a long period of time in order for a change to be noticed. The Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com...
  • Page 73: Figure 24: P-Factor Computation Example

    Wait until Cl<X One cycle of Cl adding ≤ ≤ X*1.1 Cl P factor OK Figure 24: P-factor computation example Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 74: Setting Pump Period

    For this reason the process of getting a good control loop can take 15 minutes in very small systems, and up to a few hours in large systems. Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com...
  • Page 75: Technical Specifications

    SECURITY Bubble removal Automated internal mechanism Operation password Cell cleaning Automated internal mechanism Technician password *Optional Feature Datasheet rev. 1.6 Blue I Water Technologies | 18 Ha'melacha St, Rosh Ha'ayin 4809148, Israel Tel: 972-9-7680004 | Fax 972-9-7652331 | info@blueitechnologies.com www.blueitechnologies.com...
  • Page 76: Tel: 972-9-7680004 | Fax 972-9-7652331 | Info@Blueitechnologies.com

    Patents issued and pending at the time of this printing Disclaimer Blue I Water Technologies Ltd. does not accept any responsibility for any damage caused to its products by unauthorized personnel. Use of non-Blue I Water Technologies’ reagents and/or replacement parts will void all warranties.