Display - Keithley 707B User Manual

Hide thumbs Also See for 707B:
Table of Contents

Advertisement

Models 707B and 708B 199BSwitching Matrix User's Manual
LAN status light
The Models 707B and 708B has three status lights on the front panel .
The 1588 status light indicates 1588 operation. If this light is off, the 1588 feature is disabled or
improperly configured. If the light blinks at a one second rate, the instrument is the 1588 master. If the
instrument is a slave, the light will not blink.
The LAN status light is lit when the instrument is connected through the local area network (LAN) with
no errors. If this is not lit, the instrument is not connected through the LAN or there is a connection
problem.
If you are using the web interface, the LAN Status light blinks when you click ID.
The Power light is lit when power is applied to the instrument.

Display

This section describes the front-panel display of the Model 708B and the bottom display of the Model
707B.
During operation, the display provides information about the selected channel, channel pattern,
channel state, and errors. You can press DISPLAY to cycle between the display of the channel or
pattern, the closed channel list, or a screen message.
During setup, the display shows menu choices that you can use to configure the instrument.
During operation, the display shows the control status (local or remote) and the current channel, and
indicates if any channels are closed. An example is shown below.
The control status is shown in the upper left corner of the display. If REM is displayed, the instrument
is being controlled remotely (through GPIB, LAN, or USB). If REM is not displayed, control is through
the front panel.
By default, the top line of the display shows the slot, row, and column of the selected channel.
707B-900-01 Rev. B / June 2013
Figure 5: Display during operation
Section 2: Using the front-panel interface
2-5

Advertisement

Table of Contents
loading

This manual is also suitable for:

708b

Table of Contents