Replacing the Interconnect Board in the Sievers
1. Purpose:
To replace the interconnect board in a Sievers 500 RL TOC Analyzer.
2. Scope:
This procedure applies to all current Sievers 500 RL Analyzer configurations.
3. Safety:
This procedure requires access to electrical components of the Analyzer. Be sure to follow
ESD safety precautions, as instructed. This procedure should be performed by personnel
skilled in the maintenance of electromechanical instruments.
4. Materials:
Interconnect board replacement kit
Phillips screwdriver
5. Procedure:
5.1 If the Analyzer is taking measurements, press the Stop Analysis button.
5.2 Insert a USB flash memory drive into the port labeled USB on the side of the Analyzer and back up the
current Analyzer settings (Maintenance → Advanced → USB I/O → Save System). For more
information, see the section called "Saving System Settings" in the Analyzer's
Maintenance Manual
5.3 Remove the USB flash memory drive from the USB port and store it in a safe place.
5.4 Turn off power to the Analyzer. Also shut off the circuit breaker for the Analyzer.
5.5 Disable the flow of water to the Analyzer.
5.6 Open the Analyzer door.
5.7 Put on the ESD strap and attach the alligator clip to a metal component of the Analyzer.
5.8 Remove the system board.
5.8.1 Press down on the silver buttons on each side of the system board to release the locking pins
(see Figure 1).
5.8.2 Immediately in front of the pins are the board ejectors. Using your thumbs, simultaneously push
the ejectors out, toward the board guides. This will cause the board to begin to come forward.
5.8.3 Slide the board out of the guides and then out of the Analyzer.
GE Analytical Instruments, Inc. ©2005H
.
Page 1 of 4
6060 Spine Road
Boulder, CO 80301 USA
T 303 444 2009
F 303 444 9543
500 RL TOC Analyzer
*
Operation and
DIN 74740 Rev. A
Kit, Interconnect Board Repl., 500

Advertisement

loading

Summary of Contents for Denon DN-200AZB

  • Page 1 6060 Spine Road Boulder, CO 80301 USA T 303 444 2009 F 303 444 9543 Replacing the Interconnect Board in the Sievers 500 RL TOC Analyzer 1. Purpose: To replace the interconnect board in a Sievers 500 RL TOC Analyzer. 2.
  • Page 2 Figure 1 The 500 RL’s Electronics Boards 5.9 Remove the I/O board 5.9.1 Disconnect the UV lamp power connector from the I/O board (see Figure 2). Figure 2 The UV Power Connector 5.9.2 If there is wiring connected to the terminal blocks on the I/O board, remove the terminal blocks. Grasp each terminal block firmly and pull it straight out from the connector.
  • Page 3 Figure 3 Interconnect Board Wiring 5.10 Remove the interconnect board. 5.10.1 Remove the screws that attach the system board guides to the interconnect board. Remove the guides and set them aside in a safe place. 5.10.2 Disconnect all the wiring connected to the interconnect board, nine connections in all (reading clockwise: P1, P6, P4, P5, P7, P9, P2, P50, and RS-232).
  • Page 4 5.12 Close the Analyzer door. 5.13 Restore power to the Analyzer. 5.14 Return the Analyzer to normal operation. *Trademark of General Electric Company; may be registered in one or more countries. GE Analytical Instruments, Inc. ©2005H Page 4 of 4 DIN 74740 Rev.