InBody S10 USER'S MANUAL
Thank you for purchasing the InBody S10.
Please read this manual carefully and operate with care. Make sure to keep this manual for future reference.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the inbody s10 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Biospace inbody s10

  • Page 1 InBody S10 InBody S10 USER’S MANUAL Thank you for purchasing the InBody S10. Please read this manual carefully and operate with care. Make sure to keep this manual for future reference.
  • Page 2 Biospace is not liable for any damage caused by a failure to meet the requirements in this user’s manual.
  • Page 3 The accuracy of InBody S10 makes it a great index to assess the balance of body water levels. With touch screen feature, InBody S10 is more convenient to use. It displays past measurement results as well as accumulative graphs that show the changes in the results.
  • Page 4: Table Of Contents

    2. Options 5. Power Connection & Getting Started 33 6. Home Screen Appendix 7. Measurement condition settings 1. More about the InBody S10 8. Personal Profile 2. Specifications 9. How to Operate the Equipment 3. Customer Service Information 10. Results...
  • Page 5: How To Use This Manual

    2. Take a few moments to look at the pictures of instruction of this product to understand the configuration. 3. If you have clinical issues while using the InBody S10, please contact us using the e-mail address as shown below.
  • Page 6: Safety Information

    If the equipment is dismantled, the warranty is void, and service costs will be charged. If service is required, contact Biospace or the supplying agency. 8. Follow local governing ordinances and recycling plans regarding the disposal or recycling of device components.
  • Page 7 1. Cross Contamination Individuals with any kind of contagious disease or infection must not use or come in contact with the equipment. Please be sure to clean the equipment with appropriate disinfectant after each use. Never pour any liquid directly on the equipment, as it may leak and cause internal damage.
  • Page 8 Immunity, but also with Level A for Noise Emission by the Standard IEC60601-1- 2(EN60601-1-2), Electromagnetic Compatibility Requirements. 7. The InBody S10 has been designed, manufactured, and inspected under the full quality assurance system of Biospace. Biospace fulfills the international standardization system,...
  • Page 9 Indicators & Safety Symbols Disposal of old Electrical & Electronic Equipment (Application in the European Union and other European countries with separate collection system.) This symbol indicates that this product shall not be treated as household waste. Instead, it shall be handed over to the applicable collection point for the recycling of electrical and electronic equipment.
  • Page 10 Workplace Requirements □ Operation Environment 10 ~ 40 ℃ (50 ℉ ℉ Temperature range ~ 104 ) Relative humidity 30 ~ 75 % Atmospheric pressure range 70 ~ 106 kPa □ Storage Environment -20 ~ 60 ℃ (-4℉ ~ 140℉) Temperature range Relative humidity 10 ~ 95 % (No condensation)
  • Page 11: Chapter 1 Installation & Maintenance

    Chapter1. Installation & Maintenance 1. Contents of the box 2. Exterior & Functions 3. Installation Instructions 4. Transportation 5. Repacking 6. Maintenance VIII...
  • Page 12: Contents Of The Box

    1. Contents of the box When opening the box, make sure all of the following items are inside. A. Included Items ① InBody S10 1EA ② Quick installation guide 1EA ③ User’s manual CD 1EA ④ Adapter (12V, 3.4A) 1EA ⑤...
  • Page 13 To prevent physical shock, use Biospace’s packing material when shipping or transporting the equipment. Refer to this Chapter, Section 4: “Transportation.” If you desire to use InBody S10 and various optional equipment such as extensometers and blood pressure monitors, purchase ③SD400 (serial dispenser) separately.
  • Page 14: Exterior & Functions

    2. Exterior & Functions Individual part identification and functions with schematic sketches are provided below. Please inspect each component of the InBody S10 before installation to ensure there are no scratches or damage. A. LCD Monitor B. Electrodes Connection Unit C.
  • Page 15 ③ 9pin Serial Port, Female (RS-232C) Use to connect optional devices i.e. thermal printer or blood pressure monitor provided by Biospace. SD400(Serial Distributor) provided by Biospace is for the connection of several optional devices. ④ USB Host Port Use to interface with a USB printer or a USB storage device.
  • Page 16 Use the adapter provided by Biospace only. When you use the adapter cable, insert the adapter cable tightly into the InBody S10. Including the optional equipment, only the peripherals provided by Biospace can be connected to the InBody S10. For any inquiry about peripherals, contact Biospace.
  • Page 17: Installation Instructions

    (2) Connect the electrode cables to the InBody S10. Connect electrodes of RA(Right Arm), LA(Left Arm), RL(Right Leg), LL(Left Leg) to each corresponding part of the InBody S10. InBody S10 offers two types of electrodes, that is, the Touch Type electrode and Adhesive Type electrode.
  • Page 18 When you use the touch type electrodes When you use the adhesive type electrodes...
  • Page 19 ① Touch Type Connect black cable to the black port(V) of the electrode module, and connect red cable to the red port(I) of the electrode module. [ hand electrode ] [ foot electrode ] ② Adhesive Type Connect black cable to the black port(V) of the electrode module, and connect red cable to the red port(I) of the electrode module.
  • Page 20 Cart is sold separately. (1) Please place InBody S10 on the cart. Adjust InBody S10’s 4 screws on the rear side of the unit to the cart’s 4 grooves on the head unit. Screw them to fasten the InBody S10 onto the cart.
  • Page 21 ◆ HELP ⓛ : For left/right control Loosen the screw and adjust left/right direction. Fasten the screw again. ② : For up/down control It is written as CLOSE/OPEN. ② ② Loosen and adjust high/low degree. Fasten again. ③ : When ②...
  • Page 22 (4) Connect the electrode cables to the InBody S10. Connect electrodes of RA(Right Arm), LA(Left Arm), RL(Right Leg), LL(Left Leg) to each corresponding part of the InBody S10. InBody S10 offers two types of electrodes, that is, the Touch Type electrode and Adhesive Type electrode.
  • Page 23 When you use the adhesive type electrodes ① Touch Type Connect black cable to the black port(V) of the electrode module, and connect red cable to the red port(I) of the electrode module. [hand electrode] [foot electrode] ② Adhesive Type electrode Connect black cable to the black port(V) of the electrode module, and connect red cable to the red port(I) of the electrode module.
  • Page 24 (5) How to install thermal printer onto the cart Thermal printers are sold separately. ① Supporting metal plate is fixed on cart basket’s rear side. When two screws are unscrewed, cart and basket becomes separated. ② Adjust the position of thermal printer’s rear side to the separated supporting metal plate. Fasten 4 screws to fix them together.
  • Page 25 ④ Thermal printer cables are connected in the same manner as shown in the picture.
  • Page 26: Transportation

    If it must be transported, be extra careful to ensure safe handling. The following are some tips for safely transporting the InBody S10. (1) Before transporting the InBody S10, turn off the power switch and unplug the adapter. (2) Be careful not to damage the hand module.
  • Page 27: Repacking

    (1) Turn off the power switch. (2) Remove all cables connected to the InBody S10. (3) Put the separated units in the provided portable bag. (4) You can carry the machine simply with the portable bag.
  • Page 28: Maintenance

    (3) Do not impose severe shock or force to the machine. (4) When the InBody S10 is not in use for a period longer than one day, unplug the adapter. (5) In case of moving the equipment with the power on, take care not to make physical impact on the equipment.
  • Page 29: Chapter 2 Management & Results Description

    Chapter 2. Management & Results Description 1. Cautions Before Measurement 2. Exterior and Functions of Keypad 3. Measurement Posture 4. Connecting the electrodes 5. Power Connection & Getting Started 6. Home Screen 7. Measurement condition settings 8. Personal Profile 9. How to Operate the Equipment 10.
  • Page 30: Cautions Before Measurement

    1. Cautions Before Measurement To observe changes of the human body through body composition analysis, it is crucial to perform the analysis each time under the same conditions, temperature, posture, etc. Bear in mind, the following factors affect the results of body composition analysis, and as a result, affect the reproducibility of analysis.
  • Page 31: Exterior And Function Of Keypad

    2. Exterior and Functions of Keypad The InBody S10 has a keypad for data input. To input all information, press the relevant buttons on the keypad. Keypad is located below the LCD Monitor. It can be distinguished into 2 categories by their functions.
  • Page 32 ④ button You can check accumulated analysis results of ICE(Intracellular water), ECW(extracellular water) and TBW(total body water). ⑤ button You can check accumulated analysis results of TBW/FFM(Ratio of total body water in fat free mass). ⑥ button You can check accumulated analysis results of weight-muscle mass-percent body fat (2) Select electrodes connection type ①...
  • Page 33 The COPY button will copy all data from the equipment to a USB storage device. Insert the USB storage device into the InBody S10 and press the Copy button. Do not remove the USB storage device before the printing is complete.
  • Page 34 ⑤ button Press the ABC button when entering I.D. information into the InBody S10. A keyboard will appear on the screen. ⑥ button Use the PRINT button to print the results of the last person who did the test. This button is also used when printing the results shown on the screen.
  • Page 35 ⑥ It is use for gender input when entering personal profile. ⑦ button This button is use when data input is finished or to move on to the next item.
  • Page 36: Measurement Posture

    3. Measurement Posture Maintaining the right posture during the test is essential to achieve reliable results and accuracy. It’s important to maintain examinee’s posture for 10-15 minutes so that impedance can be measured accurately. A. Lying Posture (1) It is recommended that the examinee lying posture for about 10~15 minutes before the test, so that body water may be dispersed evenly inside the body.
  • Page 37 B. Seated Posture (1) It is recommended that the examinee lying posture for about 10~15 minutes before the test, so that body water may be dispersed evenly inside the body. (2) Make sure that back is not touching the chair. Straighten back. (3) Use a cushion to support back.
  • Page 38 Attention • Make sure that the examinee maintains a straight sitting position during the test. • It is recommended that the examinee seated posture for about 10~15 minutes. • If the chair is made of iron, make sure that the examinee’s body is not in direct contact with the iron part of the chair.
  • Page 39 The examinee should not rest his/her arm on the armrest of the chair. The examinee should not slouch. Make sure bare feet do not touch the floor. Use a mat that does not conduct electricity. Make sure thighs do not touch each other. Do not position the examinee’s legs at a 90 degree angle, but stretch them slightly to the front.
  • Page 40 C. Standing Posture (1) It is recommended that the examinee standing posture for about 10~15 minutes before the test, so that body water may be dispersed evenly inside the body. (2) Make sure bare feet do not touch the floor. Use a mat that does not conduct electricity. (3) Make sure arms do not touch the trunk part of body.
  • Page 41 Make sure arms and legs do not touch each other and arms do not touch the trunk part of body. Make sure bare feet do not touch the floor. Use a mat that does not conduct electricity. When connecting the electrodes, make sure the electrode cables are not touching the floor. Place them or hang them on a cart or another object.
  • Page 42: Connecting The Electrodes

    4. Connecting the electrodes InBody S10 offers two types of electrodes, the Touch Type electrode and the Adhesive Type electrode. Check the RA, LA, RL, LL. (RA: Right Arm, LA: Left Arm, RL: Right Leg, LL: Left Leg) Check on the electrodes parts that will come into contact with examinee’s hands and feet.
  • Page 43 (2) Foot electrodes LL: connect to Left Leg. RL: connect to Right Leg. The foot electrodes should be positioned between examinee’s anklebone and heel. Try to have the electrodes in contact with as much area as possible. The part marked in red should be positioned on the inner ankle.
  • Page 44 As EKG electrode will be in direct contact with the body of the examinee, use only the product with CE certification. We officially recommend 2330 Red Dot Resting Electrode-Tab Style of 3M and RT34 Tab-Electrodes for resting ECG of SKINTACT for the InBody S10.
  • Page 45: Power Connection & Getting Started

    When connecting peripherals (printers and other optional devices) to the InBody S10, turn on the power of peripherals before turning on the InBody S10. When turning the power off, turn off the InBody S10 first before turning off the peripherals. This process will minimize...
  • Page 46: Home Screen

    6. Home Screen InBody S10 has a touch screen function for data input. Press the buttons on the LCD screen or use the keypad buttons to enter information. The home screen of the InBody S10 has numerous options for both the user’s and the examinee’s convenience.
  • Page 47: Measurement Condition Settings

    7. Measurement condition settings A. Select electrodes type Select which electrode you want to use. You can choose either Touch Type or Adhesive Type. ① button Use when using Touch Type electrodes. ② button Use when using adhesive EKG electrodes. B.
  • Page 48: Personal Profile

    Use the DEL. button to delete the data, and enter in the correct information. 2. If you press the EXIT button during the InBody S10 test, you can reenter your data. If you press EXIT button one more time, you will be transferred to the home screen.
  • Page 49: How To Operate The Equipment

    9. How to Operate the Equipment (1) Please select electrodes type. (Touch Type, Adhesive Type) (2) Please select a examinee’s posture type. (Lying Posture, Seated Posture, Standing Posture) (3) Please select whether you will be using the ‘Dialysis Mode’ or not. (Enable, Disable) If examinees are examinees receiving kidney dialysis, make sure to set the equipment to ‘Dialysis Mode’.
  • Page 50 (9) After the test is completed, a “Test Complete” message will appear on the information screen. If you entered the examinee’s I.D. for the test, the InBody S10 will automatically save all results in the system. If the printer is connected and there are results sheets in the printer, the printer will automatically print the results on the results sheet.
  • Page 51: Results

    10. Results A. Result Screen You can check the impedance results first. If you want to check other results such as body water level, body composition, cumulative body water level, cumulative TBW/FFM, and cumulative weight/muscle/percent fat, press the relevant keypad button. Please refer to the help section of the information screen.
  • Page 52 4) ICW, ICW, ECW History 5) TBW/FFM History 6) Weight/ Muscle/PBF History...
  • Page 53 Use the printers recommended by Biospace. (2) Results Sheet Type The results sheet is consumable. Use A4 standard size paper or the printed results sheet which Biospace provides. Please contact Biospace or an authorized distributor to place the purchase order. <Body Composition Results Sheet>...
  • Page 54 If you want to create or modify user information, contact Biospace or distributor. (3) Body Composition Analysis The InBody S10 assigns a quantitative value to the select elements of the body’s composition. These values demonstrate the weight of each compositional element which comprises the examinee’s total body weight.
  • Page 55 ④ Mineral Mass (kg) Mineral Mass cannot be obtained with BIA methodology, but InBody S10 offers the estimated value of mineral mass because bone mineral mass is closely correlated with FFM. The correctness of this estimated value has been validated by comparison with the DEXA method.
  • Page 56 ① Weight (kg) Generally, BMI 18.5~25 is used for determining normal range of weight. But in InBody S10, normal range for weight is standard weight ±15% of standard value, and the range is very similar to the one based on BMI (18.5~25).
  • Page 57 Body Mass Index is a convenient way of assessing the degree of obesity. Body composition and Lean Body Mass are not considered. BMI is determined using only the weight and height. The InBody S10 identifies a standard BMI of 22 for males and 21.5 for western females and 21 for asian females.
  • Page 58 (5) Segmental Lean Analysis The number next to the bar represents the absolute value for lean body mass of an examinee in kilograms. In the range, 100% actually determines the length of the graph. It represents ideal lean body mass in the ideal weight of the examinee to his or her height. Right Arm (kg) : The value shows the amount of muscles in right arm.
  • Page 59 ① Segmental Water Analysis(Body Water) The InBody S10 is available to show the body water of each segmental body part. Body water is proportionate to soft lean mass. If soft lean mass has increased due to exercise or other efforts, the body water should have increased as well.
  • Page 60 BMR( Basal Metabolic Rate) Basal Metabolic Rate (BMR) is the minimal energy requirement for sustaining vital functions at rest. With InBody S10, BMR is estimated by a known regression equation based on FFM. FFM is known to be closely related to BMR.
  • Page 61 h. TBW/FFM The TBW/FFM ratio is a ratio of the amount of body water in the fat free mass. For a healthy individual, the TBW/FFM should be about 73.3%. While hydration level is relatively consistent for healthy individuals, it may fluctuate depending on nutritional status, infection, disease, etc.
  • Page 62: Chapter 3 Setup Establishment

    Chapter3. Setup Establishment 1. Setup 2. Setup Menu 3. DATABASE...
  • Page 63: Setup

    1. Setup The InBody S10 has a function to modify the setting for the user’s preference. The setup menu consists of Settings, Result Sheet Option, Result Output Option, Printer, Interface and Touch Alignment. Set up menu screen as shown below. If you want to save the change of the set up, press ‘ENTER’...
  • Page 64: Setup Menu

    2. SETUP Menu Items out of small categories will be listed. For selecting an option, direction button (▲,▼) appears on the right. You can change it using this direction button (▲,▼). A. Settings (1) Date / Time / Display Mode ...
  • Page 65  Body Com. Sheet Type: Select an option to print results sheet. ① Printed Paper : To use printed results sheet provided by BIOSPACE. ② Blank Paper : To use blank A4 size. All formats of the results sheet will be printed out.
  • Page 66 (1) Printer Type Select the type of printer. Printers that support PCL3 or above and SPL are compatible with the InBody S10. (PCL Compatible, SPL Compatible, SPL 2009 Compatible) (2) Result Sheet Alignment It is possible to adjust the coordinates on the results sheet. After adjustment, you can check whether the alignment has done properly by “test print”.
  • Page 67 If a related function does not operate when pressing a part of the touch screen, there is a problem with the input system. Touch Alignment enables the user to solve this problem. When you run Touch Alignment, InBody S10’s power will be turned on/off automatically. G. Example for Environmental Setup (1) To adjust sound volume ①...
  • Page 68: Database

    If you input an I.D. of personal information, the data is automatically saved. Data saved by DATABASE screen are available for search, result view, print, delete, data backup and data recovery. InBody S10 saves maximum 100,000 data. A. Data Search In DATABASE screen, all the data saved in the system will be automatically searched.
  • Page 69 When you restore data from a USB storage device, InBody S10’s existing database will be replaced by the data in the USB storage device. Please make sure that it is okay to get rid of the InBody S10’s database before using this feature.
  • Page 70: Chapter 4 Problems & Solutions

    Chapter4. Problems & Solutions 1. Error Messages 2. Troubleshooting 3. Frequently Asked Question (FAQs)
  • Page 71: Error Messages

    1. Error Messages The InBody S10 displays the following error messages to warn the user of the problems run into during operations and to guide the users to take steps. The following are the most common error messages and the steps to handle the corresponding errors.
  • Page 72: Troubleshooting

    Cause 3 When an adapter not provided by Biospace is used. Action 3 Use the adapter provided by Biospace only. Cause 4 Adapter is not tightly inserted into the InBody S10. Action 4 Insert the adapter into the power input port tightly.
  • Page 73 Cause 1 Occurs when the printer cable is unplugged. Action 1 Ensure the cable is connected tightly to the InBody S10. Occasionally this may occur as a result of a bad cable. In this case, you must replace the cable.
  • Page 74 A problem arises when the orientations of printing set in the printer doesn’t correspond with that of the InBody S10. Refer to the user’s manual of the printer to change the orientations of printing in the printer. The orientation of printing set in the InBody S10 is portrait.
  • Page 75: Frequently Asked Question (Faqs)

    C. Is the electrical current applied to a human body through electrodes safe? Yes. The BIA method uses an electrical current, but is practically harmless. The InBody S10 has acquired the CE and other certifications that assure the safety of the medical equipment.
  • Page 76 G. What does the examinee follow for accurate analysis? For accurate analysis, Biospace recommends the following: - Measure with an empty stomach. - Measure 2 hours after a meal or on an empty stomach. - Measure after urination and excretion.
  • Page 77: Chapter 5 Consumables

    Chapter5. Consumables 1. Consumables 2. Options...
  • Page 78: Consumables

    1. Consumables A. Results Sheet When using the InBody S10 with a printer, it is strongly recommended to use the results sheet supplied by Biospace. If more results sheet are needed, please contact Biospace. Results sheet size 210mm * 297mm (A4 type)
  • Page 79: Options

    The InBody S10 is inconvenient to turn on and off while carrying it around indoors. To facilitate for this inconvenience, use the IB Battery 110 with the InBody S10 and the adapter of the InBody S10. Even if the power cable of adapter is taken out, InBody S10 will not turn off.
  • Page 80 1) Function of Parts a. Connection Cable This is a cable for connecting InBody S10 and IB Battery 110. The larger end is connected to InBody S10 and the smaller end is connected to the IB Battery 110. b. Adapter Connection Socket This is a socket for connecting the adaptor to recharge the IB Battery 110.
  • Page 81 Connection Cable · Please use the connection cable for only connecting the IB Battery 110 to InBody S10. · Do not insert both sides of the connection cable to both sockets of the IB Battery 110.
  • Page 82 3) How to Recharge IB Battery 110 Recharge by inserting the adapter of InBody S10 into the adapter socket of the IB Battery 110. Please recharge using the adapter provided by InBody S10. * Adapter Specification . Manufacturer : BridgePower Corp.
  • Page 83 (3) Installing and Using InBody S10 It is recommend to connect the InBody S10 to IB Battery 110, and then use it while the adapter is connected to the IB Battery 110. a. Connect the connection cable to InBody S10 and IB Battery 110.
  • Page 84 IB Battery 110 connection diagram ▶When use InBody S10 bag ▶When use InBody S10 cart...
  • Page 85: Appendix

    Appendix 1. More about the InBody S10 2. Specifications 3. Customer Service Information...
  • Page 86: More About The Inbody S10

    Then, we can get the value of fat free mass using the volume of body water. Body fat mass is determined by deducting the lean body mass from the measured weight. Height should be entered by the user. Weight can be directly measured on the InBody S10.
  • Page 87 B. Core Technology The following are key features that make the InBody S10 extremely convenient, timely, and accurate. (1) Tetrapolar 8-Point Tactile Electrode It was a complex and inconvenient procedure to attach and detach the electrodes to a specific spot every time. Trained technicians were needed for each measurement.
  • Page 88: Specifications

    Display Type 800 X 480 Touch Color LCD External Interface RS-232C 1EA, USB Slave 1EA, USB Host 1EA Laser/inkjet PCL 3 or above and SPL(Printer recommended by BIOSPACE) Compatible Printer Thermal Printer (Optional) Dimensions 202(W) X 322(L) X 53(H): mm / 7.9(W) X 16.6(L) X 2(H) : inch Machine Weight 2kg (4.4lbs)
  • Page 89: Customer Service Information

    3. Customer Service Information Corporate agents of the InBody S10 and addresses are listed below. Contact us for assistance or more information about the InBody S10. Biospace Co., Ltd. 518-10, Dogok 2-Dong, Gangnam-gu, Seoul 135-784 KOREA TEL: 82-2-501-3939 FAX: 82-2-501-3978 Website: www.inbody.com...