REED R5002 Operator's Manual

High voltage insulation resistance tester

Advertisement

Quick Links

Model
Pantone 534 Blue - 100/80/30/5
High Voltage Insulation
Pantone 485 Red - 10/100/100/5
Pantone 123 Yellow - 0/27/100/
Resistance Tester
www
1-877-849-2127 | info@reedinstruments.com
www.reedinstruments.com
R5002
reedinstruments
Black
com

Advertisement

Table of Contents
loading

Summary of Contents for REED R5002

  • Page 1 Black R5002 Model Pantone 534 Blue - 100/80/30/5 High Voltage Insulation Pantone 485 Red - 10/100/100/5 Pantone 123 Yellow - 0/27/100/ Resistance Tester reedinstruments 1-877-849-2127 | info@reedinstruments.com www.reedinstruments.com...
  • Page 2: Table Of Contents

    Table of Contents Safety .................... 3 Features ..................4 Specifications ................4-6 Instrument Description .............7-10 Operating Instructions ............11-18 AC/DC Voltage Measurement ........... 11 Internal Power Source Measurement ........12 Low Resistance Measurement & Continuity Check ....12 Measuring Insulation Resistance ..........13 Auto-discharge .................
  • Page 3: Safety

    Safety • Complies with IEC61010 safety measurement requirement in pol- lution degree 2, overvoltage category (CAT. IV 600V) and double insulation • Do not apply more than 600V to this meter • Do not use this meter around explosive gas, vapor, or dust •...
  • Page 4: Features

    Features • Insulation test range: 0.1MΩ to 60GΩ • Insulation test voltages: 500V, 1000V, 1500V, 5000V • AC/DC voltage: 0.5V to 600V • 200mA Continuity • Resistance: 0.1Ω to 6kΩ • Auto-discharge function & voltage output warning function • Backlight level selector •...
  • Page 5 Includes: One black alligator test lead, one green alligator test lead, one red alligator test lead, one red 2-prong alligator test lead, batteries, and a tool box DC/AC Voltage Measurement Range: 0.5-600.0V Resolution: 0.1V Accuracy: AC: +1.0% reading + 5 digits (40 to 60Hz) +2.5% reading + 10 digits (61 to 400Hz) DC: +1.0% reading + 5 digits Insulation Resistance Measurement...
  • Page 6 Blue - 100/80/30/5 Pantone 534 Blue Yellow - 0/27/100/0 Pantone 123 Yellow For service on this or any other REED product or information on other Red - 10/100/100/5 Pantone 485 Red Yellow - 0/27/100/0 REED products, contact REED Instruments at info@reedinstruments.com...
  • Page 7: Instrument Description

    Instrument Description 1. Insulation resistance test/lock button 2. Relative/Time-down button 3. Data hold button 4. Internal battery check/time-up button 5. MAX/MIN & Test button 6. LCD display 7. Power adaptor input terminal 8. EARTH: high resistance measurement input 9. V/Ω input terminal & GUARD: grounding protection input terminal 10.
  • Page 8: Input Terminals

    Input Terminals Insulation Resistance Measurement LINE: 500V to 5000V range insulation resistance measurement/high voltage line output terminal (LINE + COM: two plugs red test lead to one alligator clip) COM: Return terminal for all measurements (LINE + COM: two prong red test lead) GUARD: Grounding protection input terminal (black test led) EARTH: Insulation resistance measurement input terminal (green test lead)
  • Page 9 Front Panel HOLD: Freezes current reading Backlight: Turns the backlight on, makes it brighter, and turns it off Insulation Resistance Measurement TEST/LOCK: Press and hold to test or lock on insulation resistance measurement Time-up: Adjust timer up Time-down: Adjust timer down Polarization index measurement DAR: Dielectric absorption ratio measurement...
  • Page 10: Lcd Display

    LCD Display 1. Auto-ranging indicator 2. Battery life indicator 3. High voltage output or >30V AC/DC voltage indicator 4. Indicator for automatically discharged 5. Continuity buzzer indicator 6. Analogue bar graph 7. Unit symbols 8. Function indicators 9. Clock 10. Timer 1 & Timer 2 indicators 11.
  • Page 11: Operating Instructions

    Operating Instructions After pressing the TEST button under insulation measurement, the wire will generate high voltage. Do not touch it to avoid electric shock. AC/DC Voltage Measurement To avoid electric shock, do not test if voltage to ground AC/DC is more than 600V, even if the line voltage is below 600V.
  • Page 12: Internal Power Source Measurement

    Internal Power Source Measurement 1. Disconnect the test leads. 2. Set the switch to DC. 3. Press the BATT button. 4. Read the voltage value on the LCD. 5. After 5 seconds the meter will auto-off or press the BATT button to turn off the battery voltage function.
  • Page 13: Measuring Insulation Resistance

    Measuring Insulation Resistance • Prior to testing check the circuit with a high voltage detector to be sure that there is no electrical charge • It is highly recommended to wear insulated gloves while taking high voltage measurements • Do not make measurements with the battery compartment open •...
  • Page 14: Auto-Discharge

    Auto-discharge This function releases the capacitance stored in a circuit after testing. The LCD will flash “AUTO DISCHARGE” when the meter is in discharge condition and the live voltage monitor will display “O V”. This function will be released by removing test leads 2 sec or more before discharge is complete.
  • Page 15: Use Of Guard Terminal

    1. Press the PI/DAR button to select PI or DAR Measurement. 2. The LCD will display PI & Time1/Time2 or DAR & Time1/Time2. 3. To begin measuring, press and turn the “PRESS TO TEST” button clockwise till it locks. 4. After testing, turn the button to counterclockwise to the initial posi- tion.
  • Page 16: Data Hold

    This removes the surface leakage resistance of the cable insulation so you only measure the volume resistance of the insulator. Make sure to use the Guard cord supplied with this instrument to connect the instrument to the Guard terminal. Data Hold This function freezes the reading on the display.
  • Page 17: Max/Min Value

    4. Press the MAX/MIN button again, and a “MAX MIN PEAK” symbol will appear on the LCD. The meter will display the present reading, but will continue to update and store the max and min readings. 5. To exit MAX/MIN-PEAK mode press the PEAK button or hold the MAX/MIN button for 2 seconds.
  • Page 18: Emergency Stop

    Emergency Stop If you are taking high-voltage insulation resistance measurement and the danger of a short-circuit arises, stop testing immediately by pressing the “TEST” button. Backlight Press the Backlight button to select the display backlight levels and to turn the backlight on and off. The backlight will automatically turn off in 60 seconds.
  • Page 19 Notes _________________________________________ ________________________________________________ ________________________________________________ ________________________________________________ ________________________________________________ ________________________________________________ ________________________________________________ ________________________________________________ ________________________________________________ ________________________________________________ ________________________________________________ ________________________________________________ ________________________________________________ ________________________________________________ ________________________________________________ ________________________________________________ ________________________________________________ Blue - 100/80/30/5 Pantone 534 Blue Yellow - 0/27/100/0 Pantone 123 Yellow ________________________________________________ Red - 10/100/100/5 Pantone 485 Red Yellow - 0/27/100/0 Pantone 123 Yellow Blue - 100/80/30/5 Pantone 534 Blue...
  • Page 20 Notes _________________________________________ ________________________________________________ ________________________________________________ ________________________________________________ ________________________________________________ ________________________________________________ ________________________________________________ ________________________________________________ ________________________________________________ ________________________________________________ ________________________________________________ ________________________________________________ ________________________________________________ ________________________________________________ ________________________________________________ ________________________________________________ ________________________________________________ Blue - 100/80/30/5 Pantone 534 Blue Yellow - 0/27/100/0 Pantone 123 Yellow ________________________________________________ Red - 10/100/100/5 Pantone 485 Red Yellow - 0/27/100/0 Pantone 123 Yellow Blue - 100/80/30/5 Pantone 534 Blue...