PN-X Measurement Receiver
N5264A
If neither function properly:
• One or both IP addresses could be wrong.
• One or both LAN states could be set to off.
• The crossover cable could be miswired.
• One or both analyzers could be defective.
If possible, eliminate the possibility of a defective analyzer by substitution of a known working unit. Once the
analyzer has been proven to be working properly, concentration can be placed on the network itself to
determine the cause of the failure.
Constructing a Crossover Cable
A crossover cable can be made from a standard LAN cable by connecting pin 1 from each connector to pin 3
of the other connector, and pin 2 from each connector to pin 6 of the other connector.
1. Strip away a few inches of the outside jacket insulation from the middle of a standard LAN cable that has
an RJ-45 connector on each end.
NOTE
Pins 1, 2, 3, and 6 of the connectors must be located to determine which wires to cut in the
following steps. Most, but not all, LAN cables use the color coding listed in
cable does not use this color scheme, you will have to determine the locations of the
appropriate wires before proceeding with this procedure.
Service Guide N5264-90001
Troubleshooting
Rear Panel Troubleshooting
Table
4-3. If your