Omega Engineering OM5-C User Manual page 42

Isolated signal conditioners
Table of Contents

Advertisement

SPECIFICATIONS
Typical at T
A
Module
Input
Signal Range
Standard Frequency Range
Extended Frequency Range
Impedance
Coupling
Protection
(1)
Continuous
Transient
Output
Signal Range
Current Limit
Voltage Limit
Resistance
Protection
Ripple and Noise (100Khz)
Accuracy
(2)(3)
Sinusoid
50/60 Hz
45Hz to 1kHz
1kHz to 20kHz
Non-Sinusoid
Crest Factor = 1 to 2
Crest Factor = 2 to 3
Crest Factor = 3 to 4
Crest Factor = 4 to 5
Vs. Temperature
Isolation (Common Mode)
Input to Output, Input to Power
Continuous
Transient
Output to Power
Continuous
Rejection (50-60Hz Common Mode)
Response Time (0 to 99%)
Output Enable Control
Selection Time
Voltage
Max Logic "0"
Min / Max Logic "1"
Current, "0,1"
Loop Voltage
Load Resistance (maximum)
Supply Voltage
Current
Sensitivity
Environmental
Operating Temp. Range
Storage Temp. Range
Relative Humidity
Emissions
EN50081-1, ISM Group 1, Class A (Radiated, Conducted)
Immunity
EN50082-1, ISM Group 1, Class A (ESD, RF, EFT)
Dimensions
NOTE:
(1) Module rating only. Backpanel connector rating may differ — use the lowest of the two in
accordance with required safety requirements.
(2) At standard 60Hz factory calibration. Consult factory for calibration at other frequencies.
(3) For 10-100% rated span. Add an additional 0.25% error for 0-10% Span measurements.
Accuracy includes nonlinearity, hysteresis and repeatability but not source or external shunt
inaccuracy (if used).
= +25 C and +5V power.
OM5-IAC
100mV to 10V rms, 0 to 5A rms
45Hz to 1000Hz
1kHz to 20kHz
1 M
1% shunted by 100pF (-01 thru -05),
0.05
1% (-06, -07)
AC
480V (Peak AC & DC)
ANSI/IEEE C37.90.1-1989
0-5V
8mA
18V
50
Continuous Short to Ground
0.025% Span rms
0.25% Span
0.25% Reading Additional Factor
0.75% Reading Additional Factor
0.05% Reading Additional Error
0.15% Reading Additional Error
0.30% Reading Additional Error
0.40% Reading Additional Error
100ppm/ C
1500Vrms max
ANSI/IEEE C37.90.1-1989
50Vdc max
100dB
<400ms
6.0 S @ C
= 0 to 2000pF
LOAD
+0.8V
+2.4V / +36V
0.5 A
+7.5Vdc min, +26Vdc max
(Loop Voltage - 14) / (Loop Current)
+5VDC 5%
30mA
200ppm/%
-40˚ C to +85˚ C
-40˚ C to +85˚ C
0 to 90% noncondensing
2.28" x 2.26" x 0.60" (58mm x 57mm x 15mm)
ORDERING INFORMATION
MODEL
INPUT* (rms)
OM5-IAC-100B-C
0-100mV
OM5-IAC-1B-C
0-1V
OM5-IAC-10B-C
0-10V
* Modules can be ordered with other input/output ranges. Consult factory for
ordering details and specifications.
9
OUTPUT* (dc)
0-5V
0-5V
0-5V

Advertisement

Table of Contents
loading

Table of Contents