12.5 Specifications of Optional Functions
FAIL/Chart End Detection and Output (/F1)
Item
FAIL output
Relay operation
Chart end detection output Stops recording and outputs a relay signal when the remaining amount of chart paper is approx.
Relay operation
Contact rating
Clamped Input Terminal (/H2)
Item
Measuring input terminal
Non-Glare Door Glass (/H3)
Item
Door glass
Computation Function (/M1)
Item
Number of computation channels
Operation
Constants
Communication input data Digital data via communication can be used in computations other than statistical computation (TLOG)
Remote input
Computation accuracy
Computation range
Equation
Execution interval
TLOG timer
Number of timers
Timer 1 and 2 modes
Reset
Printout
Trend recording
Recording span
Unit
Recording methods
Pen
Dot printing
12-14
Specifications
Relay output on CPU error
Energized during normal operation and de-energized on CPU error. Non-hold operation.
2 cm. Also, turns ON the words CHART END on the display.
The relay is energized when the above conditions are met. Non-hold operation.
250 VAC (50/60 Hz)/3 A, 250 VDC/0.1 A (for resistor load)
Specifications
Uses clamped input terminals.
Specifications
Uses a door glass with minimal reflection.
Specifications
Pen model: 8 channels, dot model: 12 channels
General arithmetic operations: Four arithmetic operations, square root, absolute, common
Statistical operations (TLOG): Average, max., min., sum, and max. – min. for each interval
Up to 30 constants can be defined
Range: –9.9999E+29 to –1.0000E–30, 0, 1.0000E–30 to 9.9999E+29
Symbol: K01 to K30
Number of communication input data points: 8 channels (pen model), 12 channels (dot model)
Range: –9.9999E+29 to –1.0000E–30, 0, 1.0000E–30 to 9.9999E+29
Symbol: C01 to C12
Use remote input status (0/1) in equations
Number of inputs: 5
Symbol: D01 to D05
32-bit floating-point computation (excluding TLOG computation)
Computed values throughout the computation remains within approx. ±3.4×10
Defined using up to 120 characters.
Scan interval
Performs statistical calculation for each interval specified by the timer.
3 (periodic printout timer, timer 1, and timer 2).
Absolute or relative
TLOG data can be reset at each interval.
Printout at each interval is possible.
–9999999 to 99999999, decimal position arbitrary.
Up to 6 characters
Zone recording and partial expanded recording available.
Measurement channels and computation channels can be assigned arbitrarily to pens.
Trend recording of measurement and computation channels. The fastest dot printing interval is as follows:
Number of channels being trend recorded
6 or less
7 to 12
13 to 18
logarithm, exponential, power, relational operations (<, ≤ , >, ≥ , =,
≠), logical operations (AND, OR, NOT, XOR)
Fastest dot printing interval
10 s
15 s
20 s
38
.