Eclipse
with autoSat
TM
®
Personal Ambulatory Oxygen System (PAOS)
TM

User Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Eclipse 5 and is the answer not in the manual?

Questions and answers

Carla C.
March 31, 2025

I have an Eclipse 5, where the yellow light, which is constant, keeps coming on and is buzzes (beeps). I checked every thing and all seemed fine, but it keeps doing it. What do I check for next.?

1 comments:
Mr. Anderson
March 31, 2025

A constant yellow light and beeping on a SeQual Eclipse 5 indicate a flow restriction causing low or blocked flow. This may be due to a clogged filter, kinked or blocked cannula, improper humidifier setup, or blocked ventilation.

This answer is automatically generated

Keeley Stephenson
May 13, 2025

Can you use 24% venturi mask on Sequal Eclipse 5 on 2lpm continuous

Summary of Contents for SeQual Eclipse 5

  • Page 1: User Manual

    Eclipse with autoSat ® Personal Ambulatory Oxygen System (PAOS) User Manual...
  • Page 2: User Controls & System Status Indicators

    User Controls & System Status Indicators Symbol Definition Symbol Definition FAA Approved Symbol: The U.S. Federal Aviation Read user manual before operation. See user manual Administration (FAA) has approved this device for for instructions. use on-board commercial aircraft. No Smoking Icon: Do not smoke near unit. Flow Setting Indicator Warnings / ALERT (Yellow) Indicator Pulse Mode Operation...
  • Page 3: Table Of Contents

    Contents User Controls & System Status Indicators ....................2 Quick Start Guide ............................4 Indications for Use ............................7 Contraindications for Use .........................7 Safety Guidelines ............................7 Recommended Operating Environments ....................9 Operating Instructions ..........................10 First Time Eclipse Operation ........................14 User Troubleshooting Table ........................16 Cleaning, Care & Routine Maintenance ....................19 General Airline Travel Information ......................21 Power Cartridge &...
  • Page 4: Quick Start Guide

    Quick Start Guide Unpack Your Eclipse Getting to Know Your Eclipse Review all Warnings, Cautions and additional device information in the rest of this manual. Eclipse Become familiar with the key features of the Eclipse and the User Control Panel. FRONT Control Panel Handle...
  • Page 5 The Eclipse User Control Panel displays important operating Powering on the Eclipse information. To connect to AC power source: Insert the AC Power Supply cord securely into the External Power Receptacle on the right side of the device. Plug the AC power cord into a grounded AC outlet. ON/OFF Button (Green) Indicator: This button powers StEP StEP...
  • Page 6 Select Flow Delivery Mode Begin Using the Eclipse Air Inlet Filter Flow Mode Button and Indicator Press the Flow Mode button to select your desired flow mode. Pressing this button repeatedly will toggle you back and forth between continuous flow and pulse dose modes. In Continuous Flow Mode the oxygen is provided at a constant flow rate between 0.5 and 3.0 LPM.
  • Page 7: Indications For Use

    Important! Safety Guidelines WARNING: NO MODIFICATION OF THIS EqUIPMENT IS Safety Instructions are defined as follows: PERMITTED WARNING: IMPORTANT SAFETY INFORMATION FOR HAZ- WARNING: THE MANUFACTURER RECOMMENDS AN ALTER- ARDS THAT MIGHT CAUSE SERIOUS INJURY. NATE SOURCE OF SUPPLEMENTAL OXYGEN IN THE EvENT OF A POWER OUTAGE, ALARM CONDITION, OR MECHANI- CAUTION: Important information for preventing damage to CAL FAILURE.
  • Page 8 WARNING: IF YOU FEEL DISCOMFORT OR ARE EXPERIENC- WARNING: SMOKING WHILE USING OXYGEN IS THE NUM- ING A MEDICAL EMERGENCY, SEEK MEDICAL ASSISTANCE BER ONE CAUSE OF FIRE INJURIES AND RELATED DEATHS. IMMEDIATELY. YOU MUST FOLLOW THESE SAFETY WARNINGS: WARNING: OPERATING THE OXYGEN CONCENTRATOR OUTSIDE OF WARNING: DO NOT ALLOW SMOKING, CANDLES, OR OPEN FLAMES THE OPERATIONAL TEMPERATURE SPECIFICATIONS CAN LIMIT THE IN THE SAME ROOM WITH THE DEvICE OR THE OXYGEN-CARRYING...
  • Page 9: Recommended Operating Environments

    Recommended Operating Environments For proper use of the device, the following chart provides Proper Placement of the Eclipse important information concerning the recommended operating Select a location for the device that avoids the intake of smoke, environments, or operating conditions. fumes and pollutants.
  • Page 10: Operating Instructions

    Operating Instructions Before Operating AC POWER SUPPLY This users manual serves as your reference to help you operate and maintain the device. If you have any questions or concerns please call your home care provider. AC Power Supply Important! DO NOT attempt to operate the Eclipse without first reading the Safety Guidelines section of this manual.
  • Page 11 • The Eclipse 5 will run at 3.0LPM on continuous flow while being operated with the DC Supply in a car. • The Eclipse 5 will charge the battery while on DC Supply in a car while running on continuos flow at 2LPM or less.
  • Page 12 Rechargeable Power Cartridge (Battery) Initial Battery Charging The new Battery supplied with your Eclipse is not fully charged when it is shipped from the factory. Before using your Eclipse for the first time, you must first fully charge the Battery. Rechargeable Power Cartridge (Battery) With the AC Power Supply plugged in, and the Battery correctly installed in the Power Compartment of the device, allow the...
  • Page 13 Typical Battery Recharge Time Typical New Power Cartridge Operating Times Flow Rate Continuous The typical time to recharge your Battery to achieve 80% Pulse Dose Mode capacity from a fully discharged Battery is between 2 and 5 (12 BPM) Setting Flow (LPM) SETTING hours, depending upon the device flow setting.
  • Page 14: First Time Eclipse Operation

    First Time Eclipse Operation Step 3: Connect the Oxygen Supply Tubing or Nasal Step 1: Positioning Your Eclipse for Use Cannula to the Oxygen Outlet Place the Eclipse in a well-ventilated, well Air Inlet Filter lit area. Be sure the air inlet and exhaust Replace the cannula and oxygen supply tubing regularly, as recommended by vents are not obstructed.
  • Page 15 Step 6: Begin Using Your Eclipse Pulse Dose Mode and autoSat Features Your Eclipse has a feature, called autoSAT that delivers a ® Breathe normally through the nasal cannula. consistent-sized bolus of oxygen up to 40 breaths per minute (see Max Breath Rate table below).
  • Page 16: User Troubleshooting Table

    User Troubleshooting Table Do Not Ignore Alarms WARNING: THE MANUFACTURER RECOMMENDS AN ALTERNATE SOURCE OF SUPPLEMENTAL OXYGEN IN THE EvENT OF A POWER OUT- AGE, ALARM CONDITION, OR MECHANICAL FAILURE. CONSULT YOUR PHYSICIAN OR EqUIPMENT PROvIDER FOR THE TYPE OF RESERvE SYSTEM REqUIRED.
  • Page 17 Red Light Symptom Possible Cause Your Action 1) Ensure the air intake filter is not clogged or restricted. Clean and replace the filter if necessary. 2) Ensure the Eclipse is in a well ventilated area. Make sure there are at least 3 inches The red light is flashing.
  • Page 18 Other Alarm Conditions Symptom Possible Cause Your Action 1) Pulse mode may not work correctly if the user has any condition that causes blocked/restricted nasal passages. (Illness, deviated septum, etc.) While in pulse mode, The Eclipse is unable 2) Ensure cannula is no longer than 7 feet in length. the compressor speeds to detect the user’s up and the pulse mode...
  • Page 19: Cleaning, Care & Routine Maintenance

    Cleaning, Care & Routine Maintenance Routine Maintenance CAUTION: Do not operate the Oxygen Concentrator with- out the air intake filter in place. If a second filter is pro- Servicing of the internal components inside the cabinet of the vided, insert the “replacement” filter before you clean the Eclipse must be conducted by a CAIRE trained and qualified dirty filter.
  • Page 20 CENTRATOR AND THE BATTERY FROM GETTING WET OR ALLOWING FLUIDS TO ENTER THE UNIT. THIS CAN CAUSE A MALFUNCTION OR life and the highest level of performance. The SeQual Battery is SHUT DOWN, AND CAUSE AN INCREASED RISK FOR ELECTRICAL the only approved Battery recommended for use with the Eclipse.
  • Page 21: General Airline Travel Information

    General Airline Travel Information Power Cartridge & Airline Travel Table During taxi, take-off, and landing, the Eclipse must be turned These tables calculate flight times only. off and stowed under the seat or in another approved stowage Please allow for ground time (preflight check location so as to not block the aisle way or the entry way into the in, security check in, etc.) and layovers.
  • Page 22 Salter Labs is a registered mark of Salter Labs, Arvin, CA 92303 Chart SeQual Technologies Inc. reserves the right to discontinue its products, or change the prices, materials, equipment, ® quality, descriptions, specifications and/or processes to its products at any time without prior notice and with no further Lysol is a registered trademark of Reckitt Benckiser, UK.