This manual is written for stereomicroscope.To ensure the safety, obtain optimum
performance and to familiarize yourself fully with the use of this microscope, it is
recommended strongly that you study this manual thoroughly before using the microscope.

Advertisement

Table of Contents
loading

Summary of Contents for M2 PZ-4

  • Page 1: User Manual

    PZ-4 STEREO MICROSCOPE User Manual M2 Associates Inc 888-402-2778 This manual is written for stereomicroscope.To ensure the safety, obtain optimum performance and to familiarize yourself fully with the use of this microscope, it is recommended strongly that you study this manual thoroughly before using the microscope.
  • Page 2: Table Of Contents

    CONTENTS USER NOTICE ................2 1. Components ................4 2. Assembly ..................5 2.1 Assembly diagram .............. 5 2.2 Detailed Assembly Procedure ........... 6 3. Adjustment and Operation ............9 3.1 Adjusting binocular head ........... 9 3.2 Focus adjustment .............. 10 4.Specification ................
  • Page 3: User Notice

    USER NOTICE 1. Safety Precaution 1. Carefully open the box, avoid the accessories, like lens, dropping to ground and being damaged. 2. Check the input voltage, be sure the input voltage which signed on the microscope is consistent with the power supply voltage, or it will bring a serious damage to the instrument. 3.
  • Page 4 If the microscope is splashed by liquid during using, cut off the power at once, and wipe up the moisture. Do not disassemble any parts of the microscope, which will affect the function or decline the performance of the microscope. Place the instrument in a cool, dry position.
  • Page 5: Components

    1. Components Binocular viewing head Adjustable WF10×eyepieces Focusing Assembly Clamping Knob Zoom body Objectives Power cord Glass stage Brightness adjusting knob Standard base Main Switch Stereo Microscope Coarse focusing knob Fine focusing knob Stereo Microscope...
  • Page 6: Assembly

    2. Assembly 2.1 Assembly diagram The diagram below shows how to assemble the various modules. The numbers indicate the order of assembly. ★When assembling the microscope, make sure that all parts are free of dust and dirt, and avoid scratching any parts or touching glass surfaces. Binocular viewing head ⑤...
  • Page 7: Detailed Assembly Procedure

    2.2 Detailed Assembly Procedure 2.2.1 Mounting binocular viewing head (Fig.1 Fig.2) ② 1. Rotate the objective① into the objective mount thread on the lower end of the zoom body. ② , as ③ in fig.2. 2. Aligning the positioning pin of the zoom body③ to ①...
  • Page 8 2.2.2 Mounting binocular head (Fig.3) Loose the clamping bolt⑦ on the pillar⑥ and insert the binocular head and zoom body assembly⑧into the bracket of the pillar, then screw down the bolt⑦. 2.2.3 Mounting eyepieces (Fig.3 Fig.4 ⑩ Fig.5) 1. Remove the eyepiece dust caps⑨ and gently insert two adjustable eyepieces ⑩...
  • Page 9 ① 2.2.6 Mounting aperture iris diaphragm attachment (Fig.8 Fig.9) ② ○Aperture iris diaphragm attachment is optional. 1. Use the provided hexagonal driver to loose the clamping bolt and Fig.8 remove the binocular viewing head ①. Mounting aperture iris diaphragm attachment② to the top end of the zoom body.
  • Page 10: Adjustment And Operation

    3. Adjustment and Operation 3.1 Adjusting binocular viewing head 3.1.1Adjusting interpupillary distance (Fig.10) Different users have different interpupillary distances. So when users change, do the interpupillary distance adjustment. While looking through the eyepieces, hold the left and right of the binocular assembly①and adjust the eyepieces by opening or closing them for binocular vision until the left and right fields of view coincide completely.
  • Page 11: Focus Adjustment

    3. 2 Focus adjustment 3.2.1 Adjusting the rotation tension of the focus (fig.13 fig.14) adjustment knob For these Microscopes ,the tension has been adjusted well before delivery. 1.Rotate the focus adjustment knob① to move the slider② to a lower place for fear that the ①...
  • Page 12: Specification

    SAFETY PRECAUTION: 1. After the equipment has been used in an observation of a specimen that is accompanied with a potential of infection, clean the parts coming in contact with the specimen to prevent infection. 2. Moving this microscope is accompanied with the risk of dropping the specimen and causing infections.
  • Page 13: Outfit

    view Dirt on eyepiece Clean eyepiece 3 . The visibility of the Dust on objective front Clean lens surface image is poor. lens. Diopter ring is not Adjust it correctly. 6. Specimen image blurs correctly adjusted. when zoom magnification is Focus specimen Not in complete focus on changed.