TENStem eco basic INSTRUCTION MANUAL

Maintenance and cleaning

No special product is required for the maintenance and cleaning
of the TENStem eco basic.
Simply clean the appliance with a soft cloth. Take care not to al-
low damp to enter the appliance. If this occurs, have the applian-
ce checked by one of our approved technicians before any further
use.

Technical check

The TENStem eco basic should be checked every 24 months.
The technical checks consist of:
1. Verification of all the appliance documents (product instruction
manual and technical sheet)
2. Verification of all the appliance equipment.
3. Verification as to any mechanical deterioration, verification of
cables and plugs/sockets.
4. Operating safety:
– Verification of the output signal with a resistance of 1 kW (cur-
rent and voltage)
– Verification of the frequency
– Verification of the impulse width.
Technical checks can only be performed by approved technicians.
These checks are set out in the product technical sheet and spe-
cify the date and name of the person who performed the checks.

Electrode compatibility and connection

The TENStem eco basic can only be used with the following elec-
trodes and cables.