excessively high or low.
Check the accuracy of the oxygen saturation and pulse rate
readings by using an appropriate calibration apparatus.
Technical Specifications
1. Dimensions: 62.0 mm (Width) × 37.0 mm (Depth) × 32.0 mm
(Height)
Weight: 42.5 g (including the height of the two AAA dry batteries)
2. Peak wavelength range of the light emitted from the probe: red light
660 nm ± 3; infrared light 905 nm ± 5.
3. Maximum optical output power of the probe: 1.2 mW for infrared light
(905 nm).
4. Normal working condition
Working Temperature
5C to 40C (41F to 104F)
Relative Humidity
15% to 80%, non-condensing
Atmospheric Pressure
70 kPa to 106 kPa
Rated Voltage
DC 3.0 V
5. Default values and conditions of alert
Parameter
Value
Upper limit: 100
Oxygen saturation
Lower limit: 94
Upper limit: 130
Pulse rate
Lower limit: 50
When the alert switch is on and the
actual measured value goes beyond
Alert condition
the preset alert parameter range, the
Oximeter gives an alert sound.
6. Technical parameters
Parameter
Value
Oxygen
35% to 100%
Display range
saturation
Pulse rate
25 bpm to 250 bpm
Oxygen
saturation
Resolution
Pulse rate
1 bpm
Oxygen
±2% (70% to 100%)
Measurement
No requirement (≤ 69%)
saturation
precision
Pulse rate
±2 bpm
Oxygen
Upper limit: 50% to 100%
saturation
Lower limit: 50% to 100%
Alert range
Upper limit: 25 bpm to 250 bpm
Pulse rate
Lower limit: 25 bpm to 250 bpm
Oxygen
± 1% of the preset value
saturation
Alert error
The greater of ±10% of the preset
Pulse rate
value and ±5 bpm
PI
Weak PI
Min. 0.2%
Safety Type
Anti-electric-shock type: internal power supply device
Anti-electric-shock degree: BF-type application part
Running mode: continuous working
Waterproof grade: IP22