SALINITY
Range
Resolution
Accuracy
Calibration
SEAWATER SIGMA
Range
Resolution
Accuracy
Calibration
TURBIDITY
Range
Resolution
Accuracy
Calibration
ISE
Ammonium-Nitrogen
Range
Resolution
Accuracy
Calibration
Chloride
Range
Resolution
Accuracy
Calibration
±2% of reading or ±0.01 PSU whichever is greater
Based on conductivity or salinity calibration
Automatic 1, 2 or 3 points at 0, 20 and 200 FNU, or custom
0.00 to 70.00 PSU
0.01 PSU
Based on conductivity calibration
0.0 to 50.0 σ
0.1 σ
± 1σ
0.0 to 99.9 FNU;
100 to 1000 FNU
0.1 FNU from 0.0 to 99.9 FNU
1 FNU from 100 to 1000 FNU
±0.3 FNU or ±2 % of reading,
whichever is greater
0.02 to 200.0 ppm Am (as NH
0.01 ppm to 1 ppm
0.1 ppm to 200.0 ppm
±5 % of reading or 2 ppm
1 or 2 point, 10 ppm and 100 ppm
0.6 to 200.0 ppm Cl (as Cl
0.01 ppm to 1 ppm
0.1 ppm to 200.0 ppm
±5 % of reading or 2 ppm
1 or 2 point, 10 ppm and 100 ppm
14
, σ
, σ
t
0
15
, σ
, σ
t
0
15
, σ
, σ
t
0
15
4
)
-
-N)
+