Power Chair Displays a Fault on the Hand Control or Omni

The R-Net control used on this power chair
is constantly monitoring for conditions that can
cause unsafe or erratic operation. When a fault
is displayed, refer to the fault code table in this
manual for a list of corrective actions.
The following identifi es which module of the
control system has registered the problem.
• PM-Power Module (Control Module)
• JSM-Joystick Module/Omni Module
• ISM-Intelligent Seating/lighting Module
Table 1, Error Codes
Trip Text
Joystick Error
Low Battery
High Battery
M1 Brake Error
M2 Brake Error
M1 Motor Error
M2 Motor Error
Inhibit Active
Jstick Cal Error
Latched Timeout
Brake Lamp Short
Left Lamp Short
Right Lamp Short
L Ind Lamp Short
R Ind Lamp Short
L Ind Lamp Failed
R Ind Lamp Failed
SUNRISE MEDICAL
Section 5

R-net Fault Codes

Corrective Action
Ensure that the joystick is centered upon power up. If it is centered,
replace the hand control.
Recharge the batteries.
section of this manual.
Verify that the battery charger is not defective. Leave the chair on for a
few minutes to drain off the excess charge. Check the condition of the
battery charger.
Verify that the left motor is connected to the control module. Complete the
Motor and Gearbox Inspection section of this manual.
Verify that the right motor is connected to the control module. Complete
the Motor and Gearbox Inspection section of this manual.
Verify that the left motor is connected to the control module. Complete the
Motor and Gearbox Inspection section of this manual.
Verify that the right motor is connected to the control module. Complete
the Motor and Gearbox Inspection section of this manual.
Cycle power. Check all cable connections. If this does not correct the
fault, contact Sunrise Technical Service for assistance.
Calibrate the joystick. If the error is still present, the hand control may be
defective.
A latch function has exceeded it preset time.
N/A
N/A
N/A
N/A
N/A
N/A
N/A
PULSE 6 SERVICE MANUAL
Identifi ed Module
PM
Module Error
Trip Text
After charging, perform the Battery Testing
Trip Code
0506
2009
PAGE 5.1