Table 2: Location criteria
Clearance
Power requirements
Temperature:
Humidity:
Light
Air quality
HP StorageWorks 1000ux/1900ux/2300ux Optical Jukebox Setup Guide
Specification
Operating temperature
Non-operating temperature (without
media)
Operating humidity
Non-operating humidity (without
media)
Caution:
Excessive dust and
debris can damage optical
media and drives.
Description
Enough room to comfortably insert disks
in the mailslot with 2-3 inches on all
other sides to allow air circulation.
Line voltage: 100 to 240 VAC
Line frequency: 50 to 60 Hz
Power consumption:
275 W maximum
10ºC to 35ºC (50ºF to 95ºF)
-40ºC to 60ºC (-40ºF to 140ºF)
10% to 80% RH
5% to 90% RH
Avoid very bright or concentrated light
as it can interfere with the optical
sensors.
Minimal sources of particulate
contamination. Avoid areas near
frequently used doors and walkways,
stacks of supplies that collect dust,
printers that create paper dust, and
smoke-filled rooms.
Unpacking