Invacare Bora Service Manual page 96

Invacare bora/spectra xtr wheelchair service manual
Hide thumbs Also See for Bora:
Table of Contents

Advertisement

SERVICE MANUAL
· Use the Phillips screwdriver to loosen the screws (1).
· Remove the power supply chain (2).
· Use the Phillips screwdriver to loosen the screws (4).
· Additionally the sensor (5) is secured with nuts.
Loosen the nuts while holding a 6 mm wrench against.
· Remove the magnet (3).
· Remove the sensor (5).
· Loosen the four bolts (1) with a 5 mm Allen
key and a 13 mm jaw spanner and remove.
· Loosen the two bolts (2) with a 24 mm jaw
spanner and a 24 mm ring spanner and
remove.
· Lift the lifter out of the vehicle frame and
replace.
· Reassembly takes place in the reverse order.
· To complete, check all vehicle functions.
96
®