TROUBLESHOOTING

Problem
Probable cause
Nothing appears in
Batteries are drained.
the display, even
when the power is
Battery terminals are
turned on.
not in the correct
position.
The cuff does not
Battery voltage is too
inflate.
low. LOW BATTERY
MARK (
[If the batteries are drained
completely, the mark does
not appear.]
The unit does not
The cuff is not fastened
properly.
measure.
Readings are too
You moved your arm or
high or too low.
body during the
measurement.
The cuff position is not
correct.
You are using the wrong
size cuff.
Other
The value is different
from that measured at a
clinic or doctor' s office.
NOTE: If the actions described above do not solve the problem, contact
A&D Medical or Auto Control Medical at the numbers listed at the
bottom of these pages. Do not attempt to repair the device yourself.
LifeSource Health Line (Toll-Free): 1-888-726-9966
E-16
Corrective action
Replace all batteries with new
ones.
Reinstall the batteries with
negative and positive terminals
matching those indicated in the
battery compartment.
Replace all batteries with new
ones.
) blinks.
Fasten the cuff correctly.
Make sure you remain very still
and quiet during the
measurement.
Sit comfortably and still. Make
sure the cuff is at the same level
as your heart.
See Pg. E-5 "Selecting the
Correct Cuff."
See Pg. E-14 "Why Measure
Blood Pressure at Home."
Remove the batteries. Place
them back properly and try the
measurement again.
Model
. . . . . . . . . . . . . . . . . . .UA-767P (UA-767 Plus)
Type
. . . . . . . . . . . . . . . . . . . . .Oscillometric
Display
. . . . . . . . . . . . . . . . . .Digital character height
Memory
. . . . . . . . . . . . . . . . . . . .30 readings
Measurement range
. . . . . . . . .Pressure: 20 mmHg to 280 mmHg
Accuracy
. . . . . . . . . . . . . . . . .Pressure: +3 mmHg or +2%,
Pressurization
. . . . . . . . . . . .Automatic, using micropump
Depressurization
. . . . . . . . . .Automatic constant speed
Power source
. . . . . . . . . . . . .4 type "AA" (1.5 volt) alkaline
Battery life
. . . . . . . . . . . . . . .Approximately 6 months with
Operating environment
. . . .50ºF to 104ºF (10ºC to 40ºC)
Storage environment
. . . . . .-4ºF to 140ºF (-20ºC to 60ºC)
Dimensions
. . . . . . . . . . . . . . .Length: 4.3" (110 mm)
Weight
. . . . . . . . . . . . . . . . . . .10.6 oz. (300 g) without batteries
Blood pressure measurements determined by the UA-767 Plus are equivalent to
those obtained by a trained observer using the cuff/stethoscope auscultation
method within the limits prescribed by the American National Standards Institute
for electronic or automated sphygmomanometers.
In Canada (Toll-Free): 1-800-463-5414

SPECIFICATIONS

Pressure (16.0 mm)/
pulse (10.0 mm)
displayed simultaneously
Pulse: 40 pulses to 200 pulses
per minute
whichever is greater
Pulse: +5%
batteries (not included) or 120V
AC Adapter (model TB:181A)
one daily measurement
Less than 85% relative humidity
Less than 95% relative humidity
Width: 5.6" (142 mm)
Height: 2.5" (64 mm)
E-17