5-3. How to Make Measurements

WARNING
1. To avoid electrical shock hazard and/or damage to
the meter, do not measure voltages that might
exceed 500V above earth ground.
2. Before using the instrument, inspect the test leads,
connectors, and probes for cracks, breaks, or
grazes in the insulation.
DC Voltage Measurement
1. Plug the red test lead into the "V Ω mA" jack and
the black test lead into the "COM" jack.
2. Set the RANGE switch to the desired DCV position.
If the voltage to be measured is not known
beforehand, set the switch to the highest range and
reduce it until a satisfactory reading is obtained.
3. Connect the test leads to the device or circuit being
measured.
4. Turn on the power of the device or circuit being
measured. The voltage value will appear on the
digital display along with the voltage polarity.
-10-
Resistance
Range
Resolution
200Ω
0.1Ω
2kΩ
1Ω
20kΩ
10Ω
200kΩ
100Ω
2000kΩ
1kΩ
OVERLOAD PROTECTION: 15 seconds maximum 250Vrms
on all ranges.
MAXIMUM OPEN CIRCUIT VOLTAGE: 3.2V.
Audible Continuity
Range Description
Built-in buzzer sounds if resistance is
less than 100Ω.
OVERLOAD PROTECTION: 250Vrms for a maximum of 15
seconds.
TR h
FE
Range
Test Condition
NPN
10µA DC
PNP
10µA DC
DIODE TEST
Measures forward resistance of a semiconductor
junction in kΩ at max. test current of 1.5mA.
Accuracy
+(0.8% of rdg+2dgt)
+(0.8% of rdg+2dgt)
+(0.8% of rdg+2dgt)
+(0.8% of rdg+2dgt)
+(1% of rdg+2dgt)
Base Current
@3.0V DC
@3.0V DC
-7-