Scheduled Maintenance - Allpro ScanX Duo DVM Instruction Manual

Digital imaging system in-line erase
Hide thumbs Also See for ScanX Duo DVM:
Table of Contents

Advertisement

Phosphor Storage Plates (PSPs).
PSP's are subject to "wear" on the black side during normal handling and use.
They can appear scratched, while the sensitive blue or white side remains relatively
"smooth". This scratched look on the black side has absolutely no effect on the
quality of the image and should be expected under normal conditions. If the
phosphor side is scratched make sure the plates are being handled properly and
not being dragged from the ScanX tray area or other surfaces that could cause
scratching of the plate. Make sure to review the Plate Care and Preparation infor-
mation provided on page 16 of this manual.
Like all precision products, the ScanX requires a certain amount of care on a
regularly scheduled basis. A well-organized maintenance program aids dependable
equipment operation and reduces problems to a minimum. Routine checks help to
detect general overall wear, and replacement of parts can often be made before
a problem occurs. Adherence to the maintenance schedule will ensure that the
ScanX Digital Imaging System will continue performing at its best with uninter-
rupted service.
Understanding this, we have established two basic maintenance kits that will help
insure continuous operation of the ScanX Digital Imaging System. The kits and
their associated parts number along with the recommended performance schedule
are listed below.
IMPORTANT:
Service Requirement
Replace transport drive belts
and springs
Replace complete transport
belt drive belt assembly

SCHEDULED MAINTENANCE

All service requiring access to the interior of the ScanX
must be performed only by an authorized dealer service
technician with the proper training.
23
MAINTENANCE
Schedule
Kit Part No.
2 years
4 years
D1920
D1932