4. Replace tubing and tie wrap.
Time-Out Test
NOTE: For this procedure, refer to FIGURE 24.1.
1. With unit off (O), disconnect the compressor connector from the main wiring harness.
2. Plug unit in and turn power on
3. The Time-Out Failure alarm should activate within 40 seconds. Refer to SensO
Indicators on page 91 for shutdown mode.
4. Replace tubing and tie wrap.
NOTE: If any alarm fails to perform to specification, contact Invacare Technical Service.
Compressor
Compressor Connector
(Disconnect from Harness)
FIGURE 24.1 Low Pressure Test - Time-Out Test
High Pressure Test
NOTE: For this procedure, refer to FIGURE 24.2 on page 90 and FIGURE 24.3 on page 90.
• High Pressure - Occurs when pressure in product tank rises beyond a preset value,
typically 158 kPa ± 7 kPa (23 p.s.i. ± 1 p.s.i.).
• Main Valve Coil - Alarm sequence occurs when the main valve coil, connection or
P.C. board circuitry has failed.
1. With the unit running and flow set at 5 L/min., remove one orange wire from the main
valve coil.
2. The main valve coil alarm should activate within 40 seconds. Refer to SensO
Indicators on page 91 for shutdown mode (FIGURE 24.3).
3. Replace tubing and tie wrap.
NOTE: If any alarm fails to perform to specification, contact Invacare Technical Service.
Part No 1118396
.
( )
89
SECTION 24—ALARM TEST
Relief Valve (Pull Out
Stem to Simulate Leak)
2
Platinum™Oxygen
2