867B/863
Users Manual
Current Ranges
True RMS
Ranges for True
Measurement Range
RMS AC Current
(% Range to % Full
and DC Current
1
300.00 μA
5% - 100%
3000.0 μA
1
5% - 100%
30.000 mA
5% - 100%
300.00 mA
5% - 100%
3.0000A
5% - 100%
10.000A
5% - 100%
1. DC ranges available on the Fluke 867B only.
Current Accuracy [±(percent of reading + counts)]
Range
DC Current
Accuracy
20 Hz to 50 Hz
300 μA
1
0.1% + 15
3000 μA
1
0.1% + 2
2
30 mA
0.05% + 15
300 mA
0.1% + 2
3A
0.2% + 15
10A
0.2% + 2
1. Ranges available on the Fluke 867B only.
2. Fluke 863 30 mA DC range accuracy 0.1% + 15.
3. Replace counts with Average AC counts for Average Responding AC measurements. In 300 μA range, floor count
increases to 20 with battery eliminator.
8-6
Ranges for Average
Responding AC
Current
Scale)
300.0 μA
3000 μA
30.00 mA
300.0 mA
3.000A
10.00A
True RMS AC Current Accuracy (Average AC counts
50 Hz to 3 kHz
1% + 10 (4)
0.75% + 10 (4)
1% + 10 (4)
0.75% + 10 (4)
1% + 10 (4)
0.75% + 10 (4)
1% + 10 (4)
0.75% + 10 (4)
1% + 10 (4)
0.75% + 10 (4)
1% + 10 (4)
0.75% + 10 (4)
Maximum Burden
Fuse Protection
Voltage
0.03V
440 mA @ 1000V
0.3V
440 mA @ 1000V
0.03V
440 mA @ 1000V
0.3V
440 mA @ 1000V
0.1V
11A @ 1000V
0.3V
11A @ 1000V
3
)
3 kHz to 10 kHz
10 kHz to 30 kHz
2% + 20 (4)
2% + 40 (4)
2% + 20 (4)
2% + 40 (4)
2% + 20 (4)
2% + 40 (4)
2% + 20 (4)
NA
NA
NA
NA
NA