Error Indicators And Troubleshooting Tips - Omron Intelli Sense BP760CAN Instruction Manual

7 series with comfit cuff
Table of Contents

Advertisement

ERROR INDICATORS AND TROUBLESHOOTING TIPS

ERROR INDICATORS
SYMBOL
Air plug disconnected.
Arm cuff not applied correctly.
Air is leaking from the arm cuff.
Cuff over or under inflated.
The arm cuff was inflated above 299
mmHg when inflating the cuff manually.
Movement during measurement.
Clothing is interfering with the
arm cuff.
Device error.
Cuff is not applied correctly.
Batteries are worn.
CAUSE
39
CORRECTION
Insert the plug securely. Refer to "Applying the
Arm Cuff".
Apply the arm cuff correctly. Refer to "Applying
the Arm Cuff".
Replace the cuff with the new one.
Repeat measurement. Remain still and do not talk
during measurement. Refer to "Taking a
Measurement".
If "E2" appears repeatedly, inflate the cuff manually
until it is 30 to 40 mmHg above your previous
measurement result. Refer to "Taking a Measurement".
Do not inflate the cuff above 299 mmHg. Refer to
"Taking a Measurement".
Repeat measurement. Remain still and do not talk
during measurement. Refer to "Taking a
Measurement".
Remove any clothing interfering with the arm cuff.
Refer to "Applying the Arm Cuff".
Contact Customer Service.
Refer to "Applying the Arm Cuff".
Replace the four batteries.
Refer to "Battery Installation".

Advertisement

Table of Contents
loading

Table of Contents