◆ The step of running control test:
1. Preparing a test strip
Take a test strip from strip vial and insert the test strip into the meter
correctly in the front direction of the meter, in the meantime, the meter
will display the code number. Make sure the code number on the
display matches the code number on the test strip vial. If it does not
match, please set the code number on screen to match the strip vial.
When the screen shows a symbol of finger tip and drop blood, press
the left button to enter glucose control solution test mode. After the
symbol of finger tip disappears, system enters the glucose control
solution test mode. In the glucose control solution testing mode, the
results are not recorded in historical memory.
H E A -23 0
Fingertip blood test mode
2. Running the glucose control solution test
Before running the glucose control solution test, always check the
expiration date. DO NOT use control solutions if expired. Before
using the control solution, gently shake the solution bottle to make
the solution well mixed. Remove the bottle cap and wipe the tip of
the bottle with a tissue. Squeeze the bottle until a tiny drop forms
at the tip of the bottle. Put this drop on the clean surface. Touch the
drop of control solution to the tip of the reaction zone of the test strip.