BLOOD GLUCOSE

MONITORING SYSTEM

HGM-112
INSTRUCTION MANUAL

Advertisement

Table of Contents
loading

Summary of Contents for Omron HGM-112

  • Page 1: Monitoring System

    BLOOD GLUCOSE MONITORING SYSTEM HGM-112 INSTRUCTION MANUAL...
  • Page 2 IMPORTANT SAFETY PRECAUTIONS READ BEFORE USE 1. Use this device ONLY for the intended use described in this manual. 2. Do NOT use accessories which are not specified by the manufacturer. 3. Do NOT use the device if it is not working properly or if it is damaged. 4.
  • Page 3: Table Of Contents

    1.1 Introduction 1.2 Intended Use 1.3 List of Box Contents 1.4 Product and Meter Display Chapter 2 Installing and Setting up the HGM-112 meter 2.1 Inserting / replacing battery 2.2 Performing a blood glucose test Chapter 3 Finger Blood Glucose Testing 3.1 Preparation before glucose testing...
  • Page 4: Chapter 1 Hgm-112 Blood Glucose System

    User needs to apply 1μL of fresh capillary whole blood for one test, and the test result will be displayed in 5 seconds.Besides that, HGM-112 also has unique strip technology with auto coding system. Users don’t need to code the glucose meter.
  • Page 5: List Of Box Contents

    4. 10 pcs Lancets 5. Carrying Case 6. Instruction Manual 7. Log Book 8. CR2032 x 1 1.4 Product and Meter Display A.HGM-112 Blood Glucose Meter Display Battery Compartment Meter Display Strip Slot - Insert test strip here Caution: This meter will not display mmol/L.
  • Page 6 BLOOD GLUCOSE MONITORING SYSTEM-HGM-112 B.HGM-STP1A Blood Glucose Test Strip Top Edge - Apply blood sample here Confirmation Window Inserting Bar - Insert this test strip into the meter’s strip slot with this face up Warning: Touch the drop of sample to the top edge of the confirmation window of the test strip.
  • Page 7: Chapter 2 Installing And Setting Up The Hgm-112 Meter

    Even the battery power is low, HGM-112 meter still can work well: You still can use HGM-112 and the test results will still be accurate, but we strongly suggest you replace the battery as soon as possible.
  • Page 8: Performing A Blood Glucose Test

    2.2 Performing a blood glucose test A. Inserting a test strip 1. Insert the test strip into strip slot correctly; the HGM-112 meter will turn on. All LCD segments will display for 1 second. 2. Then the display will show the latest test result for 2 seconds.
  • Page 9: Chapter 3 Finger Blood Glucose Testing

    Please prepare materials you need as listed below and read this instruction manual or test strip manual to ensure the correct measuring process and results. □ HGM-112 Blood Glucose Meter □ HGM-STP1A Blood Glucose Test Strip □ Antiseptic wipes or alcohol wipes □...
  • Page 10: Fingertip Blood Measurment

    BLOOD GLUCOSE MONITORING SYSTEM-HGM-112 3.2 Fingertip blood measurement Structure: AST Cap Trigger Adjustable tip Carrier Ejector Barrel A. Install the Lancet 1. Unscrew the adjustable tip of the lancing advice.Install a new lancet into the carrier correctly and make sure it is fully seated. Then remove the protective cap and replace the adjustable tip.
  • Page 11 Warning: Sterile lancet is single use only, cannot reuse. C. Preparation of test strip 4. Insert the test strip into the strip slot and the HGM-112 meter will turn on. All LCD segments will display for 1 second. * Insert the test strip (Correctly) *Do not insert the strip from reverse side.
  • Page 12 BLOOD GLUCOSE MONITORING SYSTEM-HGM-112 D. Wash your hands and the puncture site 6. Wash hands with soap and warm water. Rinse and dry thoroughly. E. Fingertip blood sampling 7. Please massage your fingertip gently to help you obtain blood. (We suggest that you puncture different areas each time.
  • Page 13 BLOOD GLUCOSE MONITORING SYSTEM-HGM-112 Sample filled Sample too small The following items will cause inaccurate results: □ The reaction zone does not fill completely. □ Applying blood more than once. □ Applying blood to test strip after a period of time.
  • Page 14: Alternate Site Testing (Ast)

    BLOOD GLUCOSE MONITORING SYSTEM-HGM-112 * After use, press the lancet needle * Push forward the ejector to dispose. point into the protective cap. Important □ Do not use test strips more than once. □ It should not be used for testing newborns.
  • Page 15 BLOOD GLUCOSE MONITORING SYSTEM-HGM-112 DO NOT test from an alternate site: • Two hours or less after eating • After exercising • If you are sick • If you think your blood glucose is low • If you often don’t notice when your blood glucose is low •...
  • Page 16: Chapter 4 Glucose Control Solution Testing (Optional)

    8. Slightly touch the top of strip with the blood to make enough blood filled in strip; you can get the test result after 5 seconds. Warning: The AST function of HGM-112 meter only can be used for palm, not for other sites such as upper arm, forearm……etc.
  • Page 17 BLOOD GLUCOSE MONITORING SYSTEM-HGM-112 When to run the glucose control solution test □ The test strip vial is opened for long time or you think the test strips have been damaged. □ When the meter could be damage or not working properly.
  • Page 18 □ Are the test strips or control solution past the “expired date or discard date? □ Is there a HGM-112 meter defect in materials or workmanship? Repeat the control test with a new test strip. If you still have any problem,...
  • Page 19: Chapter 5 Caring For Your Strips And Meter

    BLOOD GLUCOSE MONITORING SYSTEM-HGM-112 Chapter 5 Caring for Your Strips and Meter 5.1 Caring for the strips *Wash hands before testing. *Do not use strips that have expired. Check the expiration date that is printed on the package. *Record the day opened on the test strip bottle label when you first open *Discard the bottle and any remaining test strips after 90 days from the opening date.
  • Page 20: Chapter 6 Troubleshooting

    BLOOD GLUCOSE MONITORING SYSTEM-HGM-112 Chapter 6 Troubleshooting When the meter is used incorrectly, in poor condition, or other unexpected events happen, the screen will show the error message to remind the user should notice the problem or show how to resolve them.
  • Page 21: Chapter 7 Hgm-112 Performance Characteristics

    BLOOD GLUCOSE MONITORING SYSTEM-HGM-112 Chapter 7 HGM-112 Performance Characteristics Traceability: The test result of HGM-112 blood glucose meter is plasma calibrated by chemistry analyzer, and the analyzer was calibrated with a NIST traceable glucose standard solution. Measurement range: HGM-112 meter display results between 20 ~ 600 mg/dL.
  • Page 22: Chapter 8 Specifications

    Dimension: 63.2 mm x 40.6 mm x 18.0 mm (L×W×H) Weight: 25.5 g with battery Power Supply: one CR2032 battery Storage temperature: -20 C ~ 50 C (-4 F ~ 122 F) Please protect the HGM-112 from impact Please do not disassemble the HGM-112...
  • Page 23 BLOOD GLUCOSE MONITORING SYSTEM-HGM-112 Note: The measurement unit either mg/dL or mmol/L is fixed in the meter according to the country demand. You are not able to change the unit in the meter. Warning: Disposing of Your Meter When no longer needed, remove the battery and dispose of the meter according to your local regulations.
  • Page 24 BLOOD GLUCOSE MONITORING SYSTEM-HGM-112 Lancing Device and Lancets: SteriLance Medical(Suzhou) Inc. No. 68, Litanghe Road, Xiangcheng, Suzhou, 215133, China Emergo Europe Molenstraat 15, 2513 BH The Hague, The Netherlands 0197 TEL:(31)(0) 70 345-8570 FAX:(31)(0) 70 346-7299 Blood Glucose Meter, Test Strip and Control Solution: DELBio, INC.
  • Page 25 Distributed by OMRON HEALTHCARE SINGAPORE PTE LTD. 438A Alexandra Road, #05-05/08 Alexandra Technopark, Singapore 119967 www.omron-healthcare.com.sg DELBio, INC. 3F, No.252, Shangying Road, Guishan Industrial Zone, Taoyuan County 33341, 5011556300 Taiwan 12-12 REV.00 DELTRONICS (NETHERLANDS) BV Zandsteen 15, 2132 MZ Hoofddorp, The Netherlands...