Olympus DP73 Instruction Manual page 50

Microscope digital camera
Hide thumbs Also See for DP73:
Table of Contents

Advertisement

Problem
The picture is not in focus.
The 4140 x 3096 and 2070 x 1548
images are not neat.
45
Cause
The specimen is not in focus
properly.
The aperture iris diaphragm of the
condenser is open too wide.
The field iris diaphragm is not set
properly.
Lens components of the micro-
scope are contaminated or the
cover glass on the front of the
camera is stained.
The microscope and/or camera
are subjected to vibration during
recording.
The camera is subjected to
vibration during recording.
Remedy
Adjust the focus correctly with the
fine adjustment knob.
Close the aperture iris diaphragm
a little.
Adjust the field iris diaphragm until
the image circumscribes the field
of view.
Clean the objective, camera
adapter lens, condenser and/or
window lens of the microscope,
or clean the cover glass on the
bottom of the camera head.
Acquire images in an environment
in which the microscope and
camera are not vibrated. It is effec-
tive to use an anti-vibration bench.
Acquire images in an environment
in which the microscope and
camera are not vibrated. It is effec-
tive to use an anti-vibration bench.