SECTION 4—TRANSFERRING FROM
STEP 1A
STEP 2A
Moving the Patient
STANDING SLINGS - DO NOT use the standing sling in combination with the
stand-up lift as a transport device. It is intended to transfer an individual from one
resting surface to another (such as a bed to a wheelchair). Moving a person using the
standing sling in combination with the stand-up lift over any distance is not
recommended.
During transfer, with patient suspended in a sling attached to the lift, DO NOT roll
stand-up lift over objects such as carpet, raised carpet bindings, door frames, or any
uneven surfaces or obstacles that would create an imbalance of the stand-up lift and
could cause the stand-up lift to tip over. Use push handle on the mast at all times to
push or pull the stand-up lift.
NOTE: For this procedure, refer to FIGURE 4.2.
1. Using the push handle, slowly move
the patient to the desired surface.
2. Refer to Transferring to on page 23
before lowering the patient onto the
desired surface.
Reliant™ 440
STEP 1B
STEP 2D
STEP 2C
STEP 2B
FIGURE 4.1 Lifting the Patient
WARNING
Locking Lever
Rear Swivel
Caster
NOTE: Rear swivel caster
shown in unlocked position.
FIGURE 4.2 Moving the Patient
22
STEP 2E
Push Handle
Part No 1141431