8. Slightly touch the top of strip with the blood to make enough blood
filled in strip; you can get the test result after 5 seconds.
Warning: The AST function of HGM-112 meter only can be used for palm,
Warning: Please follow the lancing device manual to execute AST blood

Chapter 4 Glucose Control Solution Testing (Optional)

Warning: If you will like to perform control solution test or have any

4.1 HGM-112 system testing

In order to provide you with an accurate test result, we suggest running a
control test to check if the meter and test strips are working properly.
When the control result is within the range on the test strip container, the
test strips and meter are working properly.
Important:
□ Use Only the OMRON glucose control solution with the HGM-112
meter and HGM-STP1A strips.
□ Mark the newly opened bottle of control solution with the date opened;
discard any unused control solution which is over three months after
opening.
□ Store the bottle at 4
refrigerate it.
□ Shake the bottle of the control solution before you use it. Please
replace new control solution if control solution deteriorates or
coagulates.
□ We suggest you use OMRON glucose control solution for glucose
control testing between 15
* Before running the test, place the HGM-112 meter, test strips and
glucose control solution at room temperature for 30 minutes to allow
them to reach the new temperature before use.
14
not for other sites such as upper arm, forearm......etc.
sampling from palm.
question, please contact your local authorized distributor for
more information.
o
o
C~30
C (39
o
C~25
o
o
F~86
F) temperature. Do not
o
o
o
C (58
F~77
F).