Troubleshooting

AvantaPure Controller Troubleshooting
Problem
ERR 1 is displayed.
ERR 3 is displayed.
Time of Day incorrect.
Possible Cause
Program settings have been corrupted.
Controller does not know the position of
the camshaft. Camshaft should be
rotating to find Home position.
Camshaft is not turning during ERR 3
display.
If camshaft is turning for more than five
minutes to find Home position:
Power failure occurred.
Press any key and reset model number.
Wait for two minutes for the controller to
return to Home position. The hourglass
should be flashing on the display
indicating the motor is running.
Check that motor is connected.
Verify that motor wire harness is
connected to motor and controller
module.
Verify that optical sensor is connected
and in place.
Verify that motor gear has engaged cam
gear.
If everything is connected, try replacing in
this order:
—Wire harness
—Motor
—Optical sensor
—Controller
Verify that optical sensor is in place and
connected to wire.
Verify that camshaft is connected
appropriately.
Verify that no dirt or rubbish is clogging
any of the cam slots.
If motor continues to rotate indefinitely,
replace the following components in this
order:
—Wire harness
—Motor
—Optical sensor
—Controller
Press SET to reset the time display.