5
Startup
Table 11
Condition
POWER light is OFF
FAULT light remains ON and COMM light is flashing
(note: other lights may also be flashing)
FAULT light remains ON and Channel B
NOT READY light is ON
FAULT light remains ON and Channel B RUN light is
ON
40
Possible causes for failure
Possible causes
Power cord is disconnected or problem
with the power supply
MIO card may be loose or there may be a
problem with the MIO card
No SIMM module in the front slot or there
may be a problem with the SIMM module
installed
Problem found with SIMM module