Instrument Memory & Volatility
MXG (Series B) & EXG Memory
MXG (Series B) & EXG Memory
The following tables describe each memory type used in the base instrument and optional baseband
generator.
Table 5-1
Base Instrument (All models and options)
Memory
Component, Type
and Size
1. Main Memory
Yes
(DRAM)
512 MByte
2. Main Memory
Yes
(Flash)
512 MByte,
partitioned as
follows:
200 MByte: Boot
(Main firmware
image, Operating
system)
50 MByte: System
(Calibration/
Configuration)
1 MByte: Secure
Storage
180 MByte: User
3. Front Panel
No
Memory
(Flash)
24 KByte
Purpose/Contents
No
Firmware operating
memory.
Yes
Factory calibration
and configuration
data
a
User file system
which includes
flatness calibration,
instrument states,
and sweep lists
Yes
Front panel keyboard
controller firmware
Data Input
Method
Operating
system
Firmware
upgrades and
user-saved
a
data
,
Operating
system
16
Location in
Sanitization
Instrument
Procedure
and
Remarks
CPU board,
Turn off
not battery
instrument
backed.
power.
Volatile
memory.
CPU board
Boot Area: no
user data
User and
Secure
Storage
Areas: see
"Erase and
Sanitize All"
on page
32.
Front Panel
None
board
required (no
user data)