Honeywell APT2000 Series User Manual page 14

2-wire toroidal conductivity transmitters
Table of Contents

Advertisement

Capabilities, Connection
The Honeywell 5000TC toroidal conductivity sensor
is used to measure low to highest conductivity val-
ues. It can be used for measurements in safe areas.
toroidal
cond. sensor
receive
send
hi
lo
lo
hi
1
2
3
4
5
Fig. 9
Conductivity measurement with Honeywell
5000TC toroidal conductivity sensor
14
temp
7
8
6
The Honeywell 5000TC toroidal
conductivity sensor may only be
used in combination with the
APT2000TC-0(H)-00 Transmitter.
For special mounting conditions of
the sensor, the cell factor can vary
between 4.0 and 4.5. Therefore the
user should perform a wet calibra-
tion of each new sensor to deter-
mine the exact cell factor.
Settings for Honeywell 5000TC toroidal conductivity
sensor
conf
1200
Pt 1000
cal
1100
4.44

Hide quick links: