Attaching The Accessories; Installing The Pump Module - Honeywell GasAlertMicro 5 User Manual

Gas detector
Table of Contents

Advertisement

Attaching the Accessories

Installing the Pump Module

The pump module can be used independently to measure gas or it can
be used with the sample probe to test for gases in confined spaces. To
measure gas using the sample probe, refer to
Probe.
Figure 6. Installing the Pump Module
Attaching the Sample
Table 14. Installing the Pump Module
Item
1
Pump module
2
Sensor filter
3
Detector
4
Machine screws (2)
To install the pump module, refer to
following procedures.
1. Deactivate the detector.
2. Remove the two machine screws and the sensor cover.
Remove the sensor filter from the sensor cover and insert it
into the pump module.
3. Attach the pump module to the detector and replace the two
machine screws.
4. If sampling in a confined space, attach the sample probe. Refer
to
Attaching the Sample
If using an auxiliary filter as an additional filtration system, refer
to
Attaching the Auxiliary
5. Activate the detector. The detector performs a pump test dur-
ing the startup self-tests. Refer to
GasAlertMicro 5/PID/IR
Attaching the Accessories
Description
Figure
6.,
Table
14., and the
Probe.
Filter.
Pump
Test.
59