Honeywell ERX User Manual page 45

Electronic pressure recorder
Table of Contents

Advertisement

The last recorded measurement of gas temperature and is scaled to the selected
temperature units at 551.
Default: -40.00
503
Case Temperature
The last recorded measurement of the air temperature inside the recorder, considered an
ambient temperature measurement.
Default: -40.00
504
Case Maximum Temperature
The highest recorded case temperature measurement since the recorder was last reset.
Default: -40.00
505
Case Minimum Temperature
The lowest recorded case temperature measurement since the recorder was last reset.
Default: 165.00
506
P1 Pressure Calibration Zero Offset
The most recent span value for P1 pressure transducer and obtained during a pressure
span calibration. A value that is normally very close to 1.0000 and recalculated on each
Pressure Span Calibration
Default = 0.000
507
P1 Previous Calibration Zero Offset
The previous P1 span factor recorded in item 506 and is updated on each calibration.
Default: 0.000
508
P1 Pressure Calibration Span Factor
The most recent span value factor for P1 pressure transducer and obtained during a
pressure span calibration. The value is normally very close to 1.0000 and recalculated on
each pressure span calibration.
Default = 1.0000
509
P1 Previous Calibration Span Factor
The previous P1 span factor recorded in item 508 and is updated on each calibration.
Default: 1.0000
510
P2 Calibration Zero Offset
The most recent span value for P2 pressure transducer and obtained during a pressure
span calibration. A value that is normally very close to 0.0000 and recalculated on each
Pressure Span Calibration
Default: 0.0000
511
P2 Previous Calibration Zero Offset
The previous P2 zero offset value recorded in item 510 and is updated on each
calibration.
Default: 0.0000
Item
Item Name & Description
512
P2 Pressure Calibration Span Factor
The most recent span value factor for P2 pressure transducer and obtained during a
pressure span calibration. The value is normally very close to 1.0000 and recalculated on
each pressure span calibration.
.
.
45