Specifications
(15 C to 35 C unless otherwise noted)

Environmental

Operating Temperature ............................ -10 C to +50 C (14 F to +122 F)
Storage
With Batteries ........................................ Per battery manufacturer's specification, not to exceed storage specification without batteries.
Without Batteries ................................... -20 C to +60 C (-4 F to +140 F)
Altitude ...................................................... 2000 m
Power Requirements ................................ 6 V dc
Batteries .................................................... 4 AA batteries (alkaline)
Battery Life ................................................ >35 hours, typical usage

Electrical and Temperature Measurement (1 year)

Function
mA Measure
0 to 24 mA
V dc Measure
0 to 30 V dc
*Temperature measure
-40 C to 150 C (-40 F to 302 F)
(RTD/Ohms)
Loop Power Supply
24 V
*Temperature measurement requires optional 720RTD Pt-100 RTD probe available as an accessory.
Range
Resolution
0.001 mA
0.001 V
0.01 C, 0.01 F
N/A
Pressure Calibrator
Accuracy
0.015 % of rdg 0.002 mA
0.015 % of rdg 0.002 V
0.015 % of rdg 0.02 ; 0.1 C ( 0.2 F)
0.25 C ( 0.45 F) combined uncertainty when using
720 RTD probe accessory
24 mA @ 24 V
Specifications
35