Agilent Technologies E1343A Service Manual page 27

75000 series b relay multiplexers
Hide thumbs Also See for E1343A:
Table of Contents

Advertisement

Chs 00 - 07 and 90
HI Measurements
Ch 00 - 07 and 90
LO Measurements
Chapter 2
1. Make Hardware Connections
Turn mainframe power OFF
Connect DMM leads as shown in Figure 2-3
Turn mainframe power ON
Figure 2-3. Ch 0-7 and 90 HI Measurements Connections
2. Measure Channel 00 HI Resistance
Send
Send
Trigger the DMM with
Send
Subtract measured value of R20 from DMM reading
Enter the result in Table 2-2 for Channel 00 HI
3. Repeat for Channels 01 - 07 HI and Channel 90 HI
Repeat steps 1 and 2 for channels 01 - 07 HI and 90 HI
Use
card # and
For Channel 90 HI, subtract combined value of R20 + R26
1. Make Hardware Connections
Turn mainframe power OFF
Connect DMM leads as shown in Figure 2-4
Turn mainframe power ON
2. Measure Channel 00 LO Resistance
Send
Send
Trigger the DMM with
Send
Subtract measured value of R21 from DMM reading
Enter the result in Table 2-2 for Channel 00 LO
to multiplexer
*RST
to close chan 00, where
CLOS (@nn00)
TRIG SGL
to open channel 00
OPEN (@nn00)
and
CLOS (@nncc)
OPEN (@nncc)
= channel # (omit leading zeroes in
cc
to multiplexer
*RST
to close chan 00, where
CLOS (@nn00)
TRIG SGL
to open channel 00
OPEN (@nn00)
= card #
nn
and note reading
, where
=
nn
)
nn
= card #
nn
and note reading
Verification Tests 25