Download Print this page

Agilent Technologies 5352B Datasheet

Agilent 5352b 40 ghz cw microwave counter

Advertisement

Quick Links

Input 1
Frequency Range: 500 MHz to 40 GHz (500 MHz to 46 GHz with Opt. 005)
Sensitivity
Full Operating Environment
500 MHz to 26.5 GHz: -25 dBm
26.5 GHz to 46 GHz: dBm = 0.741 f(GHz) -44.6
@25° C (typical)
500 MHz to 26.5 GHz: -30 dBm
26.5 GHz to 46 GHz: dBm = 0.741 f(GHz) -49.6
Maximum Input: +7 dBm
Damage Level: +25 dBm
Input 2
Frequency Range
10 Hz to 525 MHz
50 ohm: 10 MHz to 525 MHz
1 Mohm: 10 Hz to 80 MHz
Sensitivity
Full operating environment
50 ohm: 10 MHz to 525 MHz, 25 mV rms (15 mV typical @ 25° C)
1 Mohm: 10 Hz to 80 MHz, 25 mV rms (15 mV typical @ 25° C)
Maximum Input
50 ohm: + 10 dBm
1 Mohm: 1V rms
Damage Level
50 ohm or 1 Mohm dc to 5 kHz: 250 V (dc + ac peak);
>5 kHz: 5.5 V rms (+28 dBm) + 1.25 X 10^6 Vrms/freq
Connector: Replaceable fuse, type BNC (female)
Resolution: Selectable, 1 Hz to 1 MHz
Agilent 5352B 40 GHz CW Microwave
Counter
Data Sheet

Advertisement

loading

Summary of Contents for Agilent Technologies 5352B

  • Page 1 Agilent 5352B 40 GHz CW Microwave Counter Data Sheet Input 1 Frequency Range: 500 MHz to 40 GHz (500 MHz to 46 GHz with Opt. 005) Sensitivity Full Operating Environment 500 MHz to 26.5 GHz: -25 dBm 26.5 GHz to 46 GHz: dBm = 0.741 f(GHz) -44.6 @25°...
  • Page 2 High Resolution 1 Mohm mode, 1 second gate: 0.01 Hz for <100 kHz input 0.01 Hz for <1 MHz input 0.02 0.1 Hz for <10 MHz input 0.03 1 Hz for >10 MHz input Timebases Standard TCXO Aging rate: 1 X 10^-7/month Temp (0 to 50°...
  • Page 3 100 V (90 to 105 Vac rms; 47.5 to 440 Hz) 115/120 V (104/126 Vac rms; 47.5 to 440 Hz) 220 V (198 to 231 Vac rms; 47.5 to 66 Hz) 230/240 V (207 Vac rms; 47.5 to 66 Hz) Size: 425 mm W X 133 mm H X 358 mm D (16-3/4 X 5-1/4 X 14 in) Weight: 11 kg (24 lb)