Pushbutton
Press to select a different pressure unit. All units are available when the pressure sensor input is used. For
U
higher pressure module inputs, inappropriate (out-of-range) units are not available. Press
U
pressing
D
Turns pressure reading damping on and off. With damping on, the Calibrator averages several
measurements before displaying a reading. Press to confirm selection of 0% and 100% output parameters.
ENTER
Press to zero the pressure display. Vent pressure to atmosphere before you press this pushbutton. With an
Z
Absolute Pressure Module, see special instructions below.
Press to read the minimum pressure and current readings since power was turned on or the registers were
N
cleared. Press again to read the maximum pressure and current readings since power was turned on. Press
and hold for 3 seconds to clear the MIN/MAX registers.
S
Press to perform switch test.
Z
a
Press to toggle the mA display mode between mA, mA Percent, and mA Percent Error.
h
h
Press
W
normal operation.
Table 4. Pushbutton Functions
to source loop voltage.
to freeze the display. The
Description
g
symbol appears on the display. Press